Real-time discrimination of ventricular tachyarrhythmia with Fourier-transform neural network

被引:308
作者
Minami, K
Nakajima, H
Toyoshima, T
机构
[1] Medtron Japan, Dept Res & Dev, Saiwai Ku, Kawasaki, Kanagawa 2100913, Japan
[2] Teikyo Univ, Sch Med, Dept Surg 2, Itabashi Ku, Tokyo 173, Japan
关键词
defibrillator; Fourier transform; neural network; QRS complex; supraventricular; tachyarrhythmia;
D O I
10.1109/10.740880
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have developed a method to discriminate life-threatening ventricular arrhythmias by observing the QRS complex of the electrocardiogram (ECG) in each heartbeat. Changes in QRS complexes due to rhythm origination and conduction path were observed with the Fourier transform, and three kinds of rhythms were discriminated by a neural network. In this paper, the potential of our method for clinical uses and real-time detection was examined using human surface ECG's and intracardiac electrograms (EGM's), The method achieved high sensitivity and specificity (greater than or equal to 0.98) in discrimination of supraventricular rhythms from ventricular ones, We also present a hardware implementation of the algorithm on a commercial single-chip CPU.
引用
收藏
页码:179 / 185
页数:7
相关论文
共 14 条
[1]  
ALFONSO VX, 1995, IEEE ENG MED BIOL, V14, P152
[2]  
[Anonymous], 1994, NEURAL NETWORKS
[3]   ALGORITHMIC SEQUENTIAL DECISION-MAKING IN THE FREQUENCY-DOMAIN FOR LIFE THREATENING VENTRICULAR ARRHYTHMIAS AND IMITATIVE ARTIFACTS - A DIAGNOSTIC SYSTEM [J].
BARRO, S ;
RUIZ, R ;
CABELLO, D ;
MIRA, J .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (04) :320-328
[4]   External defibrillators and emergency external pacemakers [J].
Charbonnier, FM .
PROCEEDINGS OF THE IEEE, 1996, 84 (03) :487-499
[5]   RECOGNITION OF VENTRICULAR-FIBRILLATION USING NEURAL NETWORKS [J].
CLAYTON, RH ;
MURRAY, A ;
CAMPBELL, RWF .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (02) :217-220
[6]   DIFFERENTIATION OF BEATS OF VENTRICULAR AND SINUS ORIGIN USING A SELF-TRAINING NEURAL-NETWORK [J].
EVANS, SJL ;
HASTINGS, H ;
BODENHEIMER, MM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (04) :611-626
[7]   Detection algorithms in implantable cardioverter defibrillators [J].
Jenkins, JM ;
Caswell, SA .
PROCEEDINGS OF THE IEEE, 1996, 84 (03) :428-445
[8]  
MINAMI K, 1996, P COMP CARD, P545
[9]  
MINAMI K, 1997, P COMP CARD, P243
[10]  
MINAMI K, 1996, P 18 ANN INT C IEEE