Electrochemically induced pH changes resulting in protein unfolding in the ion source of an electrospray mass spectrometer

被引:87
作者
Konermann, L [1 ]
Silva, EA
Sogbein, OF
机构
[1] Univ Western Ontario, Dept Chem, London, ON N6A 5B7, Canada
[2] Fed Univ Espirito Santo, Dept Chem, BR-29060900 Vitoria, ES, Brazil
关键词
D O I
10.1021/ac010545r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The operation of an electrospray ion source in the positive ion mode involves charge-balancing oxidation reactions at the liquid/metal interface of the sprayer capillary. One of these reactions is the electrolytic oxidation of water. The protons generated in this process acidify the analyte solution within the electrospray capillary. This work explores the effects of this acidification on the electrospray ionization (ESI) mass spectrum of the protein cytochrome c (cyt c). En aqueous solution containing 40% propanol, cyt c unfolds around pH 5.6. Mass spectra recorded under these conditions, using a simple ESI series circuit, display a bimodal charge-state distribution that reflects an equilibrium mixture of folded and unfolded protein in solution. These spectra are not strongly affected by electrochemical acidification. An "external loop" is added to the ESI circuit when the metal needle of the sample injection syringe is connected to ground. The resulting circuit represents two coupled electrolytic cells that share the ESI capillary as a common anode. Under these conditions, the rate of charge-balancing oxidation reactions is dramatically increased because the ion source has to supply electrons for both, the external circuit and the ESI circuit. The analytical implications of this effect are briefly discussed. Mass spectra of cyt c recorded with the syringe needle grounded are shifted to higher charge states, indicating that electrochemical acidification has caused the protein to unfold in the ion source. The acidification can be suppressed by increasing the flow rate and lowering the electrolyte concentration of the solution and by using an electrolyte that acts as redox buffer. The observed acidification is similar for sprayer capillaries made of platinum and stainless steel. Removal of the protective oxide layer on the stainless steel surface results in effective redox buffering for a few minutes.
引用
收藏
页码:4836 / 4844
页数:9
相关论文
共 66 条
[1]  
ATKINS PW, 1998, PHYSICAL CHEM
[2]   The methanol-induced conformational transitions of β-lactoglobulin, cytochrome c, and ubiquitin at low pH:: A study by electrospray ionization mass spectrometry [J].
Babu, KR ;
Moradian, A ;
Douglas, DJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2001, 12 (03) :317-328
[3]   Methanol-induced conformations of myoglobin at pH 4.0 [J].
Babu, KR ;
Douglas, DJ .
BIOCHEMISTRY, 2000, 39 (47) :14702-14710
[4]   MECHANISM OF ELECTROSPRAY MASS-SPECTROMETRY - ELECTROSPRAY AS AN ELECTROLYSIS CELL [J].
BLADES, AT ;
IKONOMOU, MG ;
KEBARLE, P .
ANALYTICAL CHEMISTRY, 1991, 63 (19) :2109-2114
[5]   ION SPRAY INTERFACE FOR COMBINED LIQUID CHROMATOGRAPHY/ATMOSPHERIC PRESSURE IONIZATION MASS-SPECTROMETRY [J].
BRUINS, AP ;
COVEY, TR ;
HENION, JD .
ANALYTICAL CHEMISTRY, 1987, 59 (22) :2642-2646
[6]   Molten globule-like state of cytochrome c under conditions simulating those near the membrane surface [J].
Bychkova, VE ;
Dujsekina, AE ;
Klenin, SI ;
Tiktopulo, EI ;
Uversky, VN ;
Ptitsyn, OB .
BIOCHEMISTRY, 1996, 35 (19) :6058-6063
[7]   PROBING CONFORMATIONAL-CHANGES IN PROTEINS BY MASS-SPECTROMETRY [J].
CHOWDHURY, SK ;
KATTA, V ;
CHAIT, BT .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1990, 112 (24) :9012-9013
[8]  
Cole R.B., 1997, Electrospray Ionization Mass Spectrometry
[9]  
de la Mora JF, 2000, J MASS SPECTROM, V35, P939
[10]   THE CURRENT EMITTED BY HIGHLY CONDUCTING TAYLOR CONES [J].
DELAMORA, JF ;
LOSCERTALES, IG .
JOURNAL OF FLUID MECHANICS, 1994, 260 :155-184