Beat-to-beat repolarization duration measurements from ambulatory 24 hour ECG analysis: Which is the most reliable method of evaluation?

被引:1
作者
Bystricky, W [1 ]
Safer, A [1 ]
Schweizer, M [1 ]
Melcher, H [1 ]
机构
[1] Knoll AG, D-6700 Ludwigshafen, Germany
来源
COMPUTERS IN CARDIOLOGY 1999, VOL 26 | 1999年 / 26卷
关键词
D O I
10.1109/CIC.1999.825931
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The quality of repolarization duration? measurements is influenced by noise. The variability of 6 standard measures are compared on the base of 35 Holter ECGs from a practical field study in part highly disturbed. RTmax corrected by Bazett's formula was found to be the most robust parameter in the given population.
引用
收藏
页码:161 / 164
页数:4
相关论文
共 6 条
[1]  
*GETEMED, 1999, LANZ EKG SYST CARDIO
[2]  
MAIER C, 1999, HMS KIIRFK C PLUS PL
[3]  
PORTA A, 1998, MED BIOL ENG COM JAN
[4]  
QIUZHEN X, 1998, J ELECTROCARDIOLOGY, V30
[5]  
ROUSSEEUW P, 1992, L1 STAT ANAL RELATED
[6]  
1999, KOUBIOS PROGRAM SYST