A stereotactic method for the three-dimensional registration of multi-modality biologic images in animals: NMR, PET, histology, and autoradiography

被引:49
作者
Humm, JL [1 ]
Ballon, D [1 ]
Hu, YC [1 ]
Ruan, S [1 ]
Chui, C [1 ]
Tulipano, PK [1 ]
Erdi, A [1 ]
Koutcher, J [1 ]
Zakian, K [1 ]
Urano, M [1 ]
Zanzonico, P [1 ]
Mattis, C [1 ]
Dyke, J [1 ]
Chen, Y [1 ]
Harrington, P [1 ]
O'Donoghue, JA [1 ]
Ling, CC [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys & Radiol, New York, NY 10021 USA
关键词
image registration; fiduciary markers; rodent imaging;
D O I
10.1118/1.1600738
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objective of this work was to develop and then validate a stereotactic fiduciary marker system for tumor xenografts in rodents which could be used to co-register magnetic resonance imaging (MRI), PET, tissue histology, autoradiography, and measurements from physiologic probes. A Teflon(TM) fiduciary template has been designed which allows the precise insertion of small hollow Teflon rods (0.71 mm diameter) into a tumor. These rods can be visualized by MRI and PET as well as by histology and autoradiography on tissue sections. The methodology has been applied and tested on a rigid phantom, on tissue phantom material, and finally on tumor bearing mice. Image registration has been performed between the MRI and PET images for the rigid Teflon phantom and among MRI, digitized microscopy images of tissue histology, and autoradiograms for both tissue phantom and tumor-bearing mice. A registration accuracy, expressed as the average Euclidean distance between the centers of three fiduciary markers among the registered image sets, of 0.2 +/-0.06 mm was achieved between MRI and microPET image sets of a rigid Teflon phantom. The fiduciary template allows digitized tissue sections to be co-registered with three-dimensional MRI images with an average accuracy of 0.21 and 0.25 mm for the tissue phantoms and tumor xenografts, respectively. Between histology and autoradiograms, it was 0.19 and 0.21 mm for tissue phantoms and tumor xenografts, respectively. The fiduciary marker system provides a coordinate system with which to correlate information from multiple image types, on a voxel-by-voxel basis, with sub-millimeter accuracy-even among imaging modalities with widely disparate spatial resolution and in the absence of identifiable anatomic landmarks. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:2303 / 2314
页数:12
相关论文
共 10 条
[1]   Using mutual information (MI) for automated 3D registration in the pelvis and thorax region for radiotherapy treatment planning [J].
Erdi, AK ;
Hu, YC ;
Chui, CS .
MEDICAL IMAGING 2000: IMAGE PROCESSING, PTS 1 AND 2, 2000, 3979 :416-425
[2]  
HILL DL, 2001, REV PHYS MED BIOL, V46, P1
[3]  
HU Y, 1999, MED PHYS, V26, P1160
[4]   THE SPATIAL ACCURACY OF CELLULAR DOSE ESTIMATES OBTAINED FROM 3D RECONSTRUCTED SERIAL TISSUE AUTORADIOGRAPHS [J].
HUMM, JL ;
MACKLIS, RM ;
LU, XQ ;
YANG, Y ;
BUMP, K ;
BERESFORD, B ;
CHIN, LM .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (01) :163-180
[5]   Consistent landmark and intensity-based image registration [J].
Johnson, HJ ;
Christensen, GE .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (05) :450-461
[6]   Multimodality image registration by maximization of mutual information [J].
Maes, F ;
Collignon, A ;
Vandermeulen, D ;
Marchal, G ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :187-198
[7]   MR SUSCEPTIBILITY MISREGISTRATION CORRECTION [J].
SUMANAWEERA, TS ;
GLOVER, GH ;
BINFORD, TO ;
ADLER, JR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (02) :251-259
[8]  
Urano M, 2002, RADIAT RES, V158, P167, DOI 10.1667/0033-7587(2002)158[0167:MOTTOT]2.0.CO
[9]  
2
[10]  
Wells W M 3rd, 1996, Med Image Anal, V1, P35