Characterization and reduction of the transient response in steady-state MR imaging

被引:157
作者
Hargreaves, BA [1 ]
Vasanawala, SS [1 ]
Pauly, JM [1 ]
Nishimura, DG [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Magnet Resonance Syst Res Lab, Stanford, CA 94305 USA
关键词
MRI; SSFP; FISP; fast imaging; transient response;
D O I
10.1002/mrm.1170
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Refocused steady-state free precession (SSFP) imaging sequences have recently regained popularity as faster gradient hardware has allowed shorter repetition times, thereby reducing SSFP's sensitivity to off-resonance effects. Although these sequences offer fast scanning with good signal-to-noise efficiency, the "transient response," or time taken to reach a steady-state, can be long compared with the total imaging time, particularly when using 2D sequences. This results in lost imaging time and has made SSFP difficult to use for real-time and cardiac-gated applications. A linear-systems analysis of the steady-state and transient response for general periodic sequences is shown. The analysis is applied to refocused-SSFP sequences to generate a two-stage method of "catalyzing," or speeding up the progression to steady-state by first scaling, then directing the magnetization. This catalyzing method is compared with previous methods in simulations and experimentally. Although the second stage of the method exhibits some sensitivity to B, variations, our results show that the transient time can be significantly reduced, allowing imaging in a shorter total scan time. Magn Reson Med 46:149-158, 2001. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:149 / 158
页数:10
相关论文
共 26 条
[1]   ELIMINATION OF TRANSVERSE COHERENCES IN FLASH MRI [J].
CRAWLEY, AP ;
WOOD, ML ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 8 (03) :248-260
[2]  
DEIMLING M, 2000, P 8 ANN M ISMRM DENV, P1500
[3]  
Deimling M., 1994, PROC 2 ANN M SMR SAN, P495
[4]   Remember true FISP? A high SNR, near 1-second imaging method for T2-like contrast in interventional MRI at .2 T [J].
Duerk, JL ;
Lewin, JS ;
Wendt, M ;
Petersilge, C .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :203-208
[5]   PHASE AND INTENSITY ANOMALIES IN FOURIER TRANSFORM NMR [J].
FREEMAN, R ;
HILL, HDW .
JOURNAL OF MAGNETIC RESONANCE, 1971, 4 (03) :366-&
[6]   STEADY-STATE FREE PRECESSION IMAGING IN THE PRESENCE OF MOTION - APPLICATION FOR IMPROVED VISUALIZATION OF THE CEREBROSPINAL-FLUID [J].
HAACKE, EM ;
WIELOPOLSKI, PA ;
TKACH, JA ;
MODIC, MT .
RADIOLOGY, 1990, 175 (02) :545-552
[7]   RAPID FOURIER-IMAGING USING STEADY-STATE FREE PRECESSION [J].
HAWKES, RC ;
PATZ, S .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :9-23
[8]  
HEID O, 1997, P 5 ANN M ISMRM VANC, P320
[9]   VERY SLOW INPLANE FLOW WITH GRADIENT ECHO IMAGING [J].
HENKELMAN, RM ;
MCVEIGH, ER ;
CRAWLEY, AP ;
KUCHARCZYK, W .
MAGNETIC RESONANCE IMAGING, 1989, 7 (04) :383-393
[10]  
HOULT DI, 1979, J MAGN RESON, V35, P66