Deep brain stimulation lead fixation: a comparative study of the Navigus and Medtronic burr hole fixation device

被引:19
作者
Wharen, RE
Putzke, JD
Uitti, RJ
机构
[1] Mayo Clin Jacksonville, Dept Neurol, Jacksonville, FL 32224 USA
[2] Mayo Clin Jacksonville, Dept Neurosurg, Jacksonville, FL 32224 USA
关键词
basecap; deep brain stimulation; lead; fixation;
D O I
10.1016/j.clineuro.2004.11.002
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: To determine the extent of lead movement based on the type of burr hole fixation device used to secure the lead (Image-Guided Neurologics [IGN], Navigus versus Medtronic [Model 7495-51]). A randomized, blinded design of lead movement measurement was used. Methods: A clinical series of 58 individuals undergoing placement of a deep brain stimulation (DBS) system with a total of 71 operative sides were examined. Lead position was compared between 71 paired, sagital, digitized X-rays of lead position immediately before and after the lead was secured to the basecap. Lead movement was measured in a randomized, blinded fashion using the Siemens measurement tool at an 8 x magnification rate. The presurgical planned target centered at the cross hatch of the reticules on a lateral X-ray served as the reference point to determine lead movement. Results: The overall mean lead movement was significantly less using the IGN (1.9 mm), as compared to the Medtronic (3.3 mm), fixation device. Moreover, the pattern of lead movement was significantly different between the two devices. That is, the majority of measured movements using IGN device was in the superior direction (74%), whereas the opposite was true for the Medtronic device (i.e., 62% with inferior movement). Conclusion: The IGN burr hole fixation device is associated with significantly less movement when securing the lead. Probable mechanisms of movement are discussed. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:393 / 395
页数:3
相关论文
共 7 条
[1]   Chronic electrical stimulation of the ventralis intermedius nucleus of the thalamus as a treatment of movement disorders [J].
Benabid, AL ;
Pollak, P ;
Gao, DM ;
Hoffmann, D ;
Limousin, P ;
Gay, E ;
Payen, I ;
Benazzouz, A .
JOURNAL OF NEUROSURGERY, 1996, 84 (02) :203-214
[2]   Subthalamic stimulation for Parkinson's disease [J].
Benabid, AL ;
Koudsié, A ;
Benazzouz, A ;
Fraix, V ;
Ashraf, A ;
Le Bas, JF ;
Chabardes, S ;
Pollak, P .
ARCHIVES OF MEDICAL RESEARCH, 2000, 31 (03) :282-289
[3]  
KELLY PJ, 1991, TUMOR STEREOTAXIS, P51
[4]   Concepts and methods in chronic thalamic stimulation for treatment of tremor: Technique and application [J].
Krauss, JK ;
Simpson, RK ;
Ondo, WG ;
Pohle, T ;
Burgunder, JM ;
Jankovic, J .
NEUROSURGERY, 2001, 48 (03) :535-541
[5]   Thalamic stimulation for the treatment of midline tremors in essential tremor patients [J].
Obwegeser, AA ;
Uitti, RJ ;
Turk, MF ;
Strongosky, AJ ;
Wharen, RE .
NEUROLOGY, 2000, 54 (12) :2342-2344
[6]   Simultaneous thalamic deep brain stimulation and implantable cardioverter-defibrillator [J].
Obwegeser, AA ;
Uitti, RJ ;
Turk, MF ;
Wszolek, UM ;
Flipse, TR ;
Smallridge, RC ;
Witte, RJ ;
Wharen, RE .
MAYO CLINIC PROCEEDINGS, 2001, 76 (01) :87-89
[7]   Quantitative and qualitative outcome measures after thalamic deep brain stimulation to treat disabling tremors [J].
Obwegeser, AA ;
Uitti, RJ ;
Witte, RJ ;
Lucas, JA ;
Turk, MF ;
Wharen, RE .
NEUROSURGERY, 2001, 48 (02) :274-281