An outbreak of bloodstream infections arising from hemodialysis equipment

被引:21
作者
Arnow, PM
Garcia-Houchins, S
Neagle, MB
Bova, JL
Dillon, JJ
Chou, T
机构
[1] Univ Chicago Hosp, Dept Med MC5065, Infect Control Program, Chicago, IL 60637 USA
[2] Univ Chicago, Pritzker Sch Med, Chicago, IL 60637 USA
[3] Univ Chicago, Dept Med, Chicago, IL 60637 USA
关键词
D O I
10.1086/515363
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
An outbreak of 29 cases of bloodstream infection by 16 pathogens occurred during 8 months at two chronic hemodialysis centers. Consequences included 21 hospital admissions and removal of 23 dialysis catheters. An epidemiologic investigation comparing case-patients with uninfected controls showed that risk was significantly (P <.05) associated with having a catheter for vascular access; receiving treatment on a Monday, Wednesday, Friday schedule; and receiving treatment on one heavily contaminated dialysis machine. Culture studies and mock trials showed that bloodstream pathogens were present in a recently installed, commercially marketed attachment for disposal of spent priming saline and could enter blood line tubing directly or indirectly during dialyzer priming and tubing assembly. The outbreak was halted by measures directed at the attachment. Investigation of this problem demonstrated that microbial overgrowth in the attachment caused bloodstream infections and underscores the importance of microbiologic considerations in the design and use of hemodialysis equipment.
引用
收藏
页码:783 / 791
页数:9
相关论文
共 27 条
[1]  
Armitage P., 1971, STAT METHODS MED RES
[2]  
*ASS ADV MED INSTR, 1995, AAMI STAND REC PRACT, V3, P27
[3]   OUTBREAK OF GRAM-NEGATIVE BACTEREMIA AND PYROGENIC REACTIONS IN A HEMODIALYSIS CENTER [J].
BECKSAGUE, CM ;
JARVIS, WR ;
BLAND, LA ;
ARDUINO, MJ ;
AGUERO, SM ;
VEROSIC, G .
AMERICAN JOURNAL OF NEPHROLOGY, 1990, 10 (05) :397-403
[4]   INFECTIONS WITH MYCOBACTERIUM-CHELONEI IN PATIENTS RECEIVING DIALYSIS AND USING PROCESSED HEMODIALYZERS [J].
BOLAN, G ;
REINGOLD, AL ;
CARSON, LA ;
SILCOX, VA ;
WOODLEY, CL ;
HAYES, PS ;
HIGHTOWER, AW ;
MCFARLAND, L ;
BROWN, JW ;
PETERSEN, NJ ;
FAVERO, MS ;
GOOD, RC ;
BROOME, CV .
JOURNAL OF INFECTIOUS DISEASES, 1985, 152 (05) :1013-1019
[6]   A PROSPECTIVE-STUDY OF THE MECHANISMS OF INFECTION ASSOCIATED WITH HEMODIALYSIS CATHETERS [J].
CHEESBROUGH, JS ;
FINCH, RG ;
BURDEN, RP .
JOURNAL OF INFECTIOUS DISEASES, 1986, 154 (04) :579-596
[7]   SERRATIA-MARCESCENS BACTEREMIA FROM CONTAMINATED PRESSURE TRANSDUCERS [J].
DONOWITZ, LG ;
MARSIK, FJ ;
HOYT, JW ;
WENZEL, RP .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1979, 242 (16) :1749-1751
[8]   FACTORS THAT INFLUENCE MICROBIAL CONTAMINATION OF FLUIDS ASSOCIATED WITH HEMODIALYSIS MACHINES [J].
FAVERO, MS ;
CARSON, LA ;
BOND, WW ;
PETERSEN, NJ .
APPLIED MICROBIOLOGY, 1974, 28 (05) :822-830
[9]  
FAVERO MS, 1996, HOSP EPIDEMIOLOGY IN, P693
[10]  
Feldman HI, 1996, J AM SOC NEPHROL, V7, P523