Effect of visible light on photo-stimulated-phosphor imaging plates

被引:6
作者
Molteni, R [1 ]
机构
[1] Dentsply Gendex, Gendex Dent Xray Div, Des Plaines, IL 60018 USA
来源
CARS 2003: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS | 2003年 / 1256卷
关键词
storage phosphor; imaging plates; photo-stimulated phosphors; photo-stimulable-phosphor radiography;
D O I
10.1016/S0531-5131(03)00345-5
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
When irradiated with X-ray radiation, photo-stimulated-phosphor (PSP) imaging plates (IP) charge up with a latent radiographic image. Such image is discharged, hence (partially) erased, on exposure to visible light. In this work, the effects of exposure to light are quantitatively investigated and reported, in order to provide proper guidelines for (a) handling radiographically exposed imaging plates prior to scanning, so as to prevent excessive fading of the latent image and loss of information due to ambient light, and (b) removing the remaining latent image prior to utilising the imaging plate for a further X-ray examination. Data are reported for three types of light and two types of photo-stimulable phosphor. (C) 2003 Published by Elsevier Science B.V.
引用
收藏
页码:1199 / 1205
页数:7
相关论文
共 5 条
[1]   Quantitative dental radiography with a new photostimulable phosphor system [J].
Couture, RA ;
Hildebolt, C .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2000, 89 (04) :498-508
[2]  
*DENTSPL GEND, DEN DIG IM SYST US M
[3]  
HALL CJ, 1992, REV SCI INSTRUM, V53, P697
[4]  
SHAW CC, 1992, P SOC PHOTO-OPT INS, V1651, P156, DOI 10.1117/12.59393
[5]  
TAKAHASHO K, 2002, TECHNICAL REV