A Multiple Local Models Approach to Accuracy Improvement in Continuous Glucose Monitoring

被引:23
作者
Barcelo-Rico, Fatima [1 ]
Bondia, Jorge [1 ]
Luis Diez, Jose [1 ]
Rossetti, Paolo [1 ]
机构
[1] Univ Politecn Valencia, Univ Inst Control & Ind Informat, Valencia, Spain
关键词
PLASMA-GLUCOSE; INTERSTITIAL GLUCOSE; CLINICAL ACCURACY; BLOOD-GLUCOSE; SENSOR; MICRODIALYSIS; HYPOGLYCEMIA; INSULIN; HUMANS; TISSUE;
D O I
10.1089/dia.2011.0138
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Background: Continuous glucose monitoring (CGM) devices estimate plasma glucose (PG) from measurements in compartments alternative to blood. The accuracy of currently available CGM is yet unsatisfactory and may depend on the implemented calibration algorithms, which do not compensate adequately for the differences of glucose dynamics between the compartments. Here we propose and validate an innovative calibration algorithm for the improvement of CGM performance. Methods: CGM data from GlucoDay (R) (A. Menarini, Florence, Italy) and paired reference PG have been obtained from eight subjects without diabetes during eu-, hypo-, and hyperglycemic hyperinsulinemic clamps. A calibration algorithm based on a dynamic global model (GM) of the relationship between PG and CGM in the interstitial space has been obtained. The GM is composed by independent local models (LMs) weighted and added. LMs are defined by a combination of inputs from the CGM and by a validity function, so that each LM represents to a variable extent a different metabolic condition and/or sensor-subject interaction. The inputs best suited for glucose estimation were the sensor current I and glucose estimations (G) over cap, at different time instants [I-k, Ik-1, (G) over cap (k-1)] (IIG). In addition to IIG, other inputs have been used to obtain the GM, achieving different configurations of the calibration algorithm. Results: Even in its simplest configuration considering only IIG, the new calibration algorithm improved the accuracy of the estimations compared with the manufacturer's estimate: mean absolute relative difference (MARD) = 10.8 +/- 1.5% versus 14.7 +/- 5.4%, respectively (P = 0.012, by analysis of variance). When additional exogenous signals were considered, the MARD improved further (7.8 +/- 2.6%, P<0.05). Conclusions: The LM technique allows for the identification of intercompartmental glucose dynamics. Inclusion of these dynamics into the calibration algorithm improves the accuracy of PG estimations.
引用
收藏
页码:74 / 82
页数:9
相关论文
共 30 条
[1]
Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring [J].
Aussedat, B ;
Dupire-Angel, M ;
Gifford, R ;
Klein, JC ;
Wilson, GS ;
Reach, G .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2000, 278 (04) :E716-E728
[2]
New possibilistic method for discovering linear local behavior using hyper-Gaussian distributed membership function [J].
Barcelo-Rico, Fatima ;
Diez, Jose-Luis ;
Bondia, Jorge .
KNOWLEDGE AND INFORMATION SYSTEMS, 2012, 30 (02) :377-403
[3]
Bequette B Wayne, 2010, J Diabetes Sci Technol, V4, P404
[4]
Bondia J, 2011, Spanish Patent application, Patent No. [P201130811, 9201130811]
[5]
Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor [J].
Boyne, MS ;
Silver, DM ;
Kaplan, J ;
Saudek, CD .
DIABETES, 2003, 52 (11) :2790-2794
[6]
Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis [J].
Clarke, WL ;
Anderson, S ;
Farhy, L ;
Breton, M ;
Gonder-Frederick, L ;
Cox, D ;
Kovatchev, B .
DIABETES CARE, 2005, 28 (10) :2412-2417
[7]
Davey Raymond J, 2010, J Diabetes Sci Technol, V4, P1393
[8]
On the meaning and use of kurtosis [J].
DeCarlo, LT .
PSYCHOLOGICAL METHODS, 1997, 2 (03) :292-307
[9]
Enhanced Accuracy of Continuous Glucose Monitoring by Online Extended Kalman Filtering [J].
Facchinetti, Andrea ;
Sparacino, Giovanni ;
Cobelli, Claudio .
DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 (05) :353-363
[10]
Tests of glycemia in diabetes [J].
Goldstein, DE ;
Little, RR ;
Lorenz, RA ;
Malone, JI ;
Nathan, D ;
Peterson, CM ;
Sacks, DB .
DIABETES CARE, 2004, 27 (07) :1761-1773