Radiation dose reduction in CT:: Application to tissue area and volume determination

被引:16
作者
Starck, G
Lönn, L
Cederblad, Å
Alpsten, M
Sjöström, L
Ekholm, S
机构
[1] Sahlgrens Univ Hosp, MR Ctr, Div Med Phys & Med Engn, S-41345 Gothenburg, Sweden
[2] Sahlgrens Univ Hosp, MR Ctr, Dept Radiat Phys, S-41345 Gothenburg, Sweden
[3] Sahlgrens Univ Hosp, MR Ctr, Dept Radiol, S-41345 Gothenburg, Sweden
[4] Sahlgrens Univ Hosp, MR Ctr, Dept Med, S-41345 Gothenburg, Sweden
关键词
computed tomography (CT); radiation exposure; tissue characterization; volume rendering; soft tissues; CT;
D O I
10.1148/radiology.209.2.9807565
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To study the potential for radiation dose reduction in computed tomographic (CT) determination of adipose and muscle tissue area and volume, which is an application in which lower spatial resolution or higher image noise can be accepted. MATERIALS AND METHODS: Measurements in vivo (in a healthy man aged 52 years; weight, 80 kg) and in phantoms were performed with two standard clinical CT scanners. Tissue areas were determined, and measurement uncertainties were analyzed. Comparisons were made with simulated and theoretic data. RESULTS: A 25-fold reduction in radiation dose was achieved and resulted in an effective dose of 13 mu Sv per section; an accuracy of greater than 5% In area measurements was maintained. The dose savings achieved in this study were close to those that are theoretically possible. CONCLUSION: Further dose reduction to about 1 mu Sv may be possible for this application. Large dose reductions are possible in CT determinations of tissue area and volume, which makes this application available also in large randomized studies in which high-performance clinical standard scanners are used. This work demonstrates that modern CT technology may have a large potential for dose reduction in specific applications.
引用
收藏
页码:397 / 403
页数:7
相关论文
共 14 条
[1]  
BOXUM J, 1985, CT DOSE MEASUREMENTS
[2]  
BROOKS R A, 1976, Medical Physics (Woodbury), V3, P237, DOI 10.1118/1.594240
[3]  
CHOWDHURY B, 1994, INT J OBESITY, V18, P219
[4]  
deBoer H, 1996, INT J OBESITY, V20, P580
[5]   ADIPOSE-TISSUE DISTRIBUTION AND PLASMA-LIPOPROTEIN LEVELS IN OBESE WOMEN - IMPORTANCE OF INTRA-ABDOMINAL FAT [J].
DESPRES, JP ;
MOORJANI, S ;
FERLAND, M ;
TREMBLAY, A ;
LUPIEN, PJ ;
NADEAU, A ;
PINAULT, S ;
THERIAULT, G ;
BOUCHARD, C .
ARTERIOSCLEROSIS, 1989, 9 (02) :203-210
[6]  
Despres JP, 1996, HUMAN BODY COMPOSITI, P149
[7]  
Ekestubbe A, 1996, Int J Oral Maxillofac Implants, V11, P650
[8]  
HAQUE P, 1981, RADIOLOGY SCULL BRAI, V5, P4101
[9]   COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[10]  
LEITZ W, 1995, RADIAT PROT DOSIM, V57, P377