In vivo rapid magnetic field measurement and shimming using single scan differential phase mapping

被引:33
作者
Kanayama, S
Kuhara, S
Satoh, K
机构
[1] Toshiba R. and D. Center, Kawasaki
[2] Commun. and Info. Syst. Res. Labs., Toshiba R. and D. Center, Saiwai-ku, Kawasaki 210
关键词
magnetic field measurement; phase mapping; automatic shimming; echo planar imaging;
D O I
10.1002/mrm.1910360421
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In vivo shimming is a powerful tool for alleviating subject-dependent magnetic field distortions in MRI, although the usually lengthy procedure has limited its use, A magnetic field measurement method called single scan differential phase mapping that can reduce the measurement time of conventional methods by more than half when used with multiple gradient-recalled echo imaging was developed, An automatic high-order shimming using the present method was implemented in a commercial MRI system, In vivo field inhomogeneity features and the shimming performances were investigated. The method was found to be useful for improving the quality of echo-planar images.
引用
收藏
页码:637 / 642
页数:6
相关论文
共 22 条
[21]   SEPARATION OF TRUE FAT AND WATER IMAGES BY CORRECTING MAGNETIC-FIELD INHOMOGENEITY INSITU [J].
YEUNG, HN ;
KORMOS, DW .
RADIOLOGY, 1986, 159 (03) :783-786
[22]   CLINICAL MAGNETIC-SUSCEPTIBILITY MAPPING OF THE BRAIN [J].
YOUNG, IR ;
KHENIA, S ;
THOMAS, DGT ;
DAVIS, CH ;
GADIAN, DG ;
COX, IJ ;
ROSS, BD ;
BYDDER, GM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (01) :2-6