Use of studded polyflex™ stents in patients with neoplastic obstructions of the central airways

被引:40
作者
Bolliger, CT [1 ]
Breitenbuecher, A
Brutsche, M
Heitz, M
Stanzel, F
机构
[1] Univ Stellenbosch, Fac Med, Dept Internal Med, Lung Unit, ZA-7505 Tygerberg, South Africa
[2] Pulmonol Tiefenauspital, Bern, Switzerland
[3] Univ Hosp, Dept Internal Med, Div Resp, Basel, Switzerland
[4] Zentralkrankenhaus, Munich, Germany
[5] Arztehaus Rietberg, Zurich, Switzerland
关键词
lung cancer; central airway obstruction; stent; polyflex (TM);
D O I
10.1159/000075654
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: The Polyflex(TM) is a commercially available silastic airway stent with proven efficacy. Due to a smooth outer surface its anchorage in the airway wall is superficial which may lead to stent migration. Objective: To study the performance of an newer version of the Polyflex stent with a studded outer surface, which should improve anchorage. Methods: In a 5-centre international study the new stent was prospectively tested in symptomatic patients with neoplastic central airway stenosis of more that 50% of normal diameter. Insertion technique, efficacy of stent placement and stent-related complications were recorded before, 1 month and 3 months post stent placement. Results: Under general anaesthesia and rigid bronchoscopy 27 stents were inserted in 26 patients, mean age: 62 years (range: 37-83), 16 men. Diagnoses were 18 bronchogenic carcinoma, 4 oesophageal carcinoma, 2 metastases, 1 tracheal carcinoma, and 1 schwannoma. The stents were inserted in the following locations: 10 right main bronchus, 8 left main bronchus, 7 trachea, and 2 tracheo-bronchial. There was significant improvement in all functional parameters assessed from before (A), to 1 month (B) and 3 months (C) after stent placement. The measured values (mean +/- SD) were for the WHO scale: A: 2.7 (0.8), B: 1.5 (0.9), C: 1.6 (1.0); for the Karnofsky scale: A: 44 (19), B: 72 (18), C: 71 (21); for the Dyspnoea Index: A: 3.3 (0.7), B: 1.5 (0.8), C: 1.9 (1.2); for FEV1: A: 1.2 (0.5), B: 1.9 (0.6), C: 1.5 (0.5), and for FVC: A: 2.1 (0.7), B: 2.8 (0.7), C: 2.5 (1.0). Stent-related complications were 4 reversible stent obstructions by secretions, 1 migration. The observation period was mean 4.3 months (range 2 days to 23 months). Conclusion: The studded Polyflex showed excellent efficacy, was very well tolerated, and had a very low migration rate. It presents an improvement over the older smooth model and can be considered a true alternative to the most widely used silastic stent, the Dumon stent. Copyright (C) 2004 S. Karger AG, Basel.
引用
收藏
页码:83 / 87
页数:5
相关论文
共 10 条
[1]  
Bolliger CT, 2000, PROG RESPIR RES, V30, P187
[2]   Evaluation of a new self-expandable silicone stent in an experimental tracheal stenosis [J].
Bolliger, CT ;
Wyser, C ;
Wu, XR ;
Hauser, R ;
Studer, W ;
Dalquen, P ;
Perruchoud, AP .
CHEST, 1999, 115 (02) :496-501
[3]  
BOLLIGER CT, 1993, CHEST, V104, P1653, DOI 10.1378/chest.104.6.1653
[4]   An Airway Wallstent for the treatment of tracheobronchial malignancies [J].
Bolliger, CT ;
Heitz, M ;
Hauser, R ;
Probst, R ;
Perruchoud, AP .
THORAX, 1996, 51 (11) :1127-1129
[5]   A DEDICATED TRACHEOBRONCHIAL STENT [J].
DUMON, JF .
CHEST, 1990, 97 (02) :328-332
[6]  
Freitag L, 2000, PROG RESPIR RES, V30, P171
[7]  
Freitag L., 1995, Journal of Bronchology, V2, P270
[8]  
Schildge J, 2001, Pneumologie, V55, P152, DOI 10.1055/s-2001-12284
[9]   Clinical and laboratory evaluation of a new thin-walled self-expanding tracheobronchial silicone stent: Progress and pitfalls [J].
Wassermann, K ;
Koch, A ;
MullerEhmsen, J ;
Reuter, M ;
Michel, O ;
Eckel, HE .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1997, 114 (04) :527-534
[10]   Concurrent glottic and tracheal stenoses: Restoration of airway continuity in end-stage malignant disease [J].
Wassermann, K ;
Mathen, F ;
Eckel, HE .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2001, 110 (04) :349-355