Identification of peak stresses in cardiac prostheses - A comparison of two-dimensional versus three-dimensional principal stress analyses

被引:29
作者
Fontaine, AA [1 ]
Ellis, JT [1 ]
Healy, TM [1 ]
Hopmeyer, J [1 ]
Yoganathan, AP [1 ]
机构
[1] GEORGIA INST TECHNOL, CARDIOVASC FLUID MECH LAB, DEPT CHEM ENGN, ATLANTA, GA 30332 USA
关键词
D O I
10.1097/00002480-199605000-00007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study assessed the accuracy of using a two-dimensional principal stress analysis compared to a three-dimensional analysis in estimating peak turbulent stresses in complex three-dimensional flows associated with cardiac prostheses. Three-component, coincident laser Doppler anemometer measurements were obtained in steady flow downstream of three prosthetic valves: a St. Jude bileaflet, Bjork-Shiley monostrut tilting disc, and Starr-Edwards ball and cage. Two-dimensional and three-dimensional principal stress analyses were performed to identify local peak stresses. Valves with locally two-dimensional flows exhibited a 10-15% underestimation of the largest measured normal stresses compared to the three-dimensional principal stresses. In nearly all flows, measured shear stresses underestimated peak principal shear stresses by 10-100%. Differences between the two-dimensional and three-dimensional principal stress analysis were less than 10% in locally two-dimensional flows. In three-dimensional flows, the two-dimensional principal stresses typically underestimated three-dimensional values by nearly 20%. However, the agreement of the two-dimensional principal stress with the three-dimensional principal stresses was dependent upon the two velocity-components used in the two-dimensional analysis, and was observed to vary across the valve flow field because of flow structure variation. The use of a two-dimensional principal stress analysis with two-component velocity data obtained from measurements misaligned with the plane of maximum mean flow shear can underpredict maximum shear stresses by as much as 100%.
引用
收藏
页码:154 / 163
页数:10
相关论文
共 21 条
  • [1] DETERMINATION OF PRINCIPAL REYNOLDS STRESSES IN PULSATILE FLOWS AFTER ELLIPTIC FILTERING OF DISCRETE VELOCITY-MEASUREMENTS
    BALDWIN, JT
    DEUTSCH, S
    PETRIE, HL
    TARBELL, JM
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04): : 396 - 403
  • [2] Blackshear P., 1972, Chemistry of Biosurfaces, P523
  • [3] Bruss K H, 1983, Life Support Syst, V1, P3
  • [4] LASER ANEMOMETRY MEASUREMENTS OF PULSATILE FLOW PAST AORTIC-VALVE PROSTHESES
    CHANDRAN, KB
    CABELL, GN
    KHALIGHI, B
    CHEN, CJ
    [J]. JOURNAL OF BIOMECHANICS, 1983, 16 (10) : 865 - 873
  • [5] A 6 YEAR EXPERIENCE WITH THE ST-JUDE MEDICAL VALVE - HEMODYNAMIC PERFORMANCE, SURGICAL RESULTS, BIOCOMPATIBILITY AND FOLLOW-UP
    CZER, LSC
    MATLOFF, J
    CHAUX, A
    DEROBERTIS, M
    YOGANATHAN, A
    GRAY, RJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (04) : 904 - 912
  • [6] FONTAINE AA, 1994, P 2 WORLD C BIOM AMS
  • [7] HINZE JO, 1975, TURBULENCE, P327
  • [8] HUNG TC, 1976, T AM SOC ART INT ORG, V22, P285
  • [9] JACOBS ML, 1985, CIRCULATION, V72, P208
  • [10] JOHNSON PL, 1989, EXP FLUIDS, V8, P137