Comparison of fatigue life for 3 types of manual wheelchairs

被引:52
作者
Fitzgerald, SG
Cooper, RA
Boninger, ML
Rentschler, AJ
机构
[1] VA Pittsburgh Healthcare Syst, Human Engn Res Labs, Pittsburgh, PA 15206 USA
[2] Univ Pittsburgh, Dept Rehabil Sci & Technol, Pittsburgh, PA 15260 USA
[3] Univ Pittsburgh, Dept Phys Med & Rehabil, Pittsburgh, PA 15260 USA
[4] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA 15260 USA
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 2001年 / 82卷 / 10期
关键词
durable medical equipment; materials testing; reimbursement mechanisms; rehabilitation; United States;
D O I
10.1053/apmr.2001.26139
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objectives: To examine 3 types of manual wheelchairs-ultralight wheelchairs (UWs), lightweight wheelchairs (LWs), and depot wheelchairs (DWs)-and to compare the fatigue life between the wheelchair types. Design: A database of different manual wheelchairs tested according to the International Organization for Standardization (ISO). Fatigue life was determined by using standards that define methods accepted internationally using double-drum and curb-drop testing equipment. Setting: A rehabilitation engineering center. Specimens: Sixty-one manual wheelchairs: 25 DWs, 22 UWs, and 14 LWs. Main Outcome Measures: Wheelchairs were examined for differences in fatigue life based on equivalent cycles. Unique survival curves were fit and compared for each wheelchair type. Results: The UWs lasted the longest, with a mean of 309,362 equivalent cycles. The DWs faired the worst, with a mean of 117,210 equivalent cycles. The Kaplan-Meier survival curves Were significantly different (p < .001). with the UWs having the longest fatigue life. Conclusion: Fatigue life for UWs is significantly greater (p < .05) than LWs and DWs, indicating wheelchairs differ in durability.
引用
收藏
页码:1484 / 1488
页数:5
相关论文
共 15 条
[1]   Wheelchair Pushrim kinetics: Body weight and median nerve function [J].
Boninger, ML ;
Cooper, RA ;
Baldwin, MA ;
Shimada, SD ;
Koontz, A .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1999, 80 (08) :910-915
[2]   Performance of selected lightweight wheelchairs on ANSI/RESNA tests [J].
Cooper, RA ;
Gonzalez, J ;
Lawrence, B ;
Renschler, A ;
Boninger, ML ;
VanSickle, DP .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1997, 78 (10) :1138-1144
[3]   Evaluation of selected ultralight manual wheelchairs using ANSI/RESNA standards [J].
Cooper, RA ;
Boninger, ML ;
Rentschler, A .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1999, 80 (04) :462-467
[4]  
Cooper RA, 1996, J REHABIL RES DEV, V33, P45
[5]  
Cooper RA, 1993, P 16 ANN RESNA C LAS, P283
[6]  
COOPER RA, 1995, REHABILITATION ENG A
[7]  
Cooper RA, 1998, WHEELCHAIR SELECTION
[8]  
Cooper RA, 1992, P 15 ANN RESNA C 199, P210
[9]  
*HLTH CAR FIN ADM, 2001, QUEST ANSW SAN ANT D
[10]  
*HLTH CAR FIN ADM, 1998, CONT ED WORKSH UN HE