viable counts;
serial dilutions;
precision;
confidence intervals;
calibration;
D O I:
10.1016/S0168-1605(03)00046-1
中图分类号:
TS2 [食品工业];
学科分类号:
0832 ;
摘要:
The basic (inherent) precision of serial dilutions and of colony counts made from them may be reliably estimated by reference solely to the pipette-manufacturer's specifications. Such estimates do not include external sources of variation and may be regarded as minima. The quality of estimation can be improved by using information gained by laboratory ('in-house') re-calibration of the pipettes. The degree of improvement was assessed by comparison with similar series made without recalibration. It was found that improvement was minimal for colony counts but worthwhile for homogeneous solutions. (C) 2003 Elsevier Science B.V. All rights reserved.