Acute in vivo testing of a respiratory assist catheter:: Implants in calves versus sheep

被引:12
作者
Eash, HJ
Frankowski, BJ
Litwak, K
Wagner, WR
Hattler, BG
Federspiel, WJ
机构
[1] Univ Pittsburgh, McGowan Inst Regenerat Med, Pittsburgh, PA 15203 USA
[2] Univ Pittsburgh, Dept Chem Engn, Pittsburgh, PA 15203 USA
[3] Univ Pittsburgh, Dept Surg, Pittsburgh, PA 15203 USA
[4] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA 15203 USA
关键词
D O I
10.1097/01.MAT.0000074991.94234.B6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A respiratory catheter that is inserted through a peripheral vein and placed within the vena cava is being developed for CO2 removal in patients with acute exacerbations of chronic obstructive pulmonary disease (COPD). The catheter uses a rapidly pulsating balloon to enhance gas exchange. in this study, the CO2 removal performance of our catheter was assessed in acute sheep implants and compared with calf implants, primarily because sheep have cardiac outputs (CO) that are more comparable with human CO and lower than calves. Respiratory catheters (25 mL balloon, 0.17 m(2)) were inserted acutely in sheep (n = 2) and calves (n = 6) through the jugular vein and placed within the vena cava in two positions: spanning the right atrium (RA) and within the inferior vena cava (IVC). The postinsertion CO in the sheep ranged from 4.1 to 7.2 L/min compared with 6.2 to 15.5 L/min for the calves. The maximum CO2 removal rates (vCO(2)) were 297 ml/min/m(2) (calf) and 282 ml/min/m(2) (sheep) in the RA position and 240 ml/min/m(2) (calf) and 248 ml/min/m(2) (sheep) in the IVC position. The respective removal rates between animal models were not statistically different (p values > 0.05 for all data sets). The dependence of the vCO(2) on balloon pulsation was also not statistically different between the animal models.
引用
收藏
页码:370 / 377
页数:8
相关论文
共 32 条
[1]   2000 Report of the AVMA Panel on Euthanasia [J].
Beaver, BV ;
Reed, W ;
Leary, S ;
McKiernan, B ;
Bain, F ;
Schultz, R ;
Bennett, BT ;
Pascoe, P ;
Shull, E ;
Cork, LC ;
Francis-Floyd, R ;
Amass, KD ;
Johnson, R ;
Schmidt, RH ;
Underwood, W ;
Thornton, GW ;
Kohn, B .
JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2001, 218 (05) :669-696
[2]  
Berne R.M., 1997, CARDIOVASCULAR PHYSL
[3]   PERMISSIVE HYPERCAPNIA AND INTRAVASCULAR OXYGENATOR IN THE TREATMENT OF PATIENTS WITH ARDS [J].
BRUNET, F ;
MIRA, JP ;
CERF, C ;
BELGHITH, M ;
SOUBRANE, O ;
TERMIGNON, JL ;
RENAUD, B ;
FIEROBE, L ;
HAMY, I ;
MONCHI, M ;
DESLANDE, E ;
BRUSSET, A ;
DHAINAUT, JF .
ARTIFICIAL ORGANS, 1994, 18 (11) :826-832
[4]   INTRAVENACAVAL MEMBRANE-OXYGENATION AND CARBON-DIOXIDE REMOVAL IN SEVERE ACUTE RESPIRATORY-FAILURE [J].
CONRAD, SA ;
EGGERSTEDT, JM ;
GRIER, LR ;
MORRIS, VF ;
ROMERO, MD .
CHEST, 1995, 107 (06) :1689-1697
[5]  
Cooney DO., 1976, BIOMEDICAL ENG PRINC
[6]  
Cox C S Jr, 1991, ASAIO Trans, V37, pM411
[7]  
DURBIN C G JR, 1992, Respiratory Care, V37, P147
[8]   Ex vivo testing of the intravenous membrane oxygenator [J].
Federspiel, WJ ;
Golob, JF ;
Merrill, TL ;
Lund, LW ;
Bultman, JA ;
Frankowski, BJ ;
Watach, M ;
Litwak, K ;
Hattler, BG .
ASAIO JOURNAL, 2000, 46 (03) :261-267
[9]   Recent progress in engineering the Pittsburgh intravenous membrane oxygenator [J].
Federspiel, WJ ;
Hewitt, T ;
Hout, MS ;
Walters, FR ;
Lund, LW ;
Sawzik, PJ ;
Reeder, G ;
Borovetz, HS ;
Hattler, BG .
ASAIO JOURNAL, 1996, 42 (05) :M435-M442
[10]  
Federspiel WJ, 1997, ASAIO J, V43, pM725