Automated Overnight Closed-Loop Glucose Control in Young Children with Type 1 Diabetes

被引:38
作者
Elleri, Daniela [1 ,2 ]
Allen, Janet M. [1 ,2 ]
Nodale, Marianna [1 ]
Wilinska, Malgorzata E. [1 ,2 ]
Mangat, Jasdip S. [3 ]
Larsen, Anne Mette F. [1 ]
Acerini, Carlo L. [2 ]
Dunger, David B. [1 ,2 ]
Hovorka, Roman [1 ,2 ]
机构
[1] Univ Cambridge, Metab Res Labs, Inst Metab Sci, Cambridge CB2 0QQ, England
[2] Univ Cambridge, Dept Pediat, Cambridge CB2 0QQ, England
[3] Addenbrookes Hosp, Dept Med Phys & Clin Engn, Cambridge, England
基金
英国医学研究理事会;
关键词
GLYCEMIC CONTROL; HYPOGLYCEMIA; ADOLESCENTS; PREVALENCE; SYSTEM; PUMP;
D O I
10.1089/dia.2010.0176
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: We evaluated the effectiveness of automated overnight closed-loop (AOCL) insulin delivery and the influence of timing of initiation on glucose control overnight in young children with type 1 diabetes (T1D). Methods: Eight children with T1D (four boys, four girls) (mean +/- SD: 9.4 +/- 2.7 years old; body mass index, 18.3 +/- 2.3 kg/m(2); duration of diabetes, 3.9 +/- 2.5years; total daily insulin dose, 0.7 +/- 0.1 U/kg/day; glycosylated hemoglobin, 7.9 +/- 0.9%) were studied in a clinical research facility on two separate occasions. Subjects had a meal at 18: 00 (77 +/- 8 g of carbohydrate [CHO]) and snack at 21: 00 (21 +/- 6 g of CHO), both accompanied by a prandial insulin bolus. In random order, AOCL was started at 18: 00 or 21: 00 h and ran until 08: 00 h the next day. Subcutaneous continuous glucose monitoring data were fed automatically into the model predictive control algorithm. Calculated subcutaneous insulin infusion rates were sent wirelessly to an insulin pump. Plasma glucose was measured to assess closed-loop performance. Results: No rescue CHOs were administered. Time spent with plasma glucose in the target range from 3.9 to 8.0 mmol/L was 50.7% (29.0%, 72.2%), and it did not differ on the two occasions: median (interquartile range), 42% (18%, 64%) versus 58% (32%, 79%) (P = 0.161). Time when plasma glucose was above 8.0 mmol/L (42% [25%, 82%] vs. 29% [14%, 64%], P = 0.093), time below 3.9 mmol/L (0% [0%, 11%] vs. 8% [0%, 17%], P = 0.500), low blood glucose index (0.1 [0.0, 2.5] vs. 1.7 [0.4, 3.3], P = 0.380), plasma glucose at the start of AOCL (12.5 +/- 2.7 vs. 11.6 +/- 4.2 mmol/L, P = 0.562), and mean overnight plasma glucose (8.3 +/- 2.1 vs. 7.5 +/- 2.2 mmol/L, P = 0.246) were also similar. Conclusions: AOCL is feasible in young children with T1D. Comparable results were obtained when closed-loop was initiated at 18:00 or 21:00 h.
引用
收藏
页码:419 / 424
页数:6
相关论文
共 23 条
[1]   MD-Logic Artificial Pancreas System A pilot study in adults with type 1 diabetes [J].
Atlas, Eran ;
Nimri, Revital ;
Miller, Shahar ;
Grunberg, Eli A. ;
Phillip, Moshe .
DIABETES CARE, 2010, 33 (05) :1072-1076
[3]  
Bequette B Wayne, 2005, Diabetes Technol Ther, V7, P28, DOI 10.1089/dia.2005.7.28
[4]   Nocturnal hypoglycemia in children and adolescents with insulin-dependent diabetes mellitus: Prevalence and risk factors [J].
Beregszaszi, M ;
TubianaRufi, N ;
Benali, K ;
Noel, M ;
Bloch, J ;
Czernichow, P .
JOURNAL OF PEDIATRICS, 1997, 131 (01) :27-33
[5]   Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes [J].
Bergenstal, Richard M. ;
Tamborlane, William V. ;
Ahmann, Andrew ;
Buse, John B. ;
Dailey, George ;
Davis, Stephen N. ;
Joyce, Carol ;
Peoples, Tim ;
Perkins, Bruce A. ;
Welsh, John B. ;
Willi, Steven M. ;
Wood, Michael A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (04) :311-320
[6]  
Chassin Ludovic J, 2005, Diabetes Technol Ther, V7, P72, DOI 10.1089/dia.2005.7.72
[7]   Assessment and management of hypoglycemia in children and adolescents with diabetes [J].
Clarke, William ;
Jones, Timothy ;
Rewers, Arleta ;
Dunger, David ;
Klingensmith, Georgeanna J. .
PEDIATRIC DIABETES, 2009, 10 :134-145
[8]   Hypoglycemia: Incidence and clinical predictors in a large population-based sample of children and adolescents with IDDM [J].
Davis, EA ;
Keating, B ;
Byrne, GC ;
Russell, M ;
Jones, TW .
DIABETES CARE, 1997, 20 (01) :22-25
[9]   Continuing stability of center differences in pediatric diabetes care:: Do advances in diabetes treatment improve outcome?: The Hvidoere study group on childhood diabetes [J].
de Beaufort, Carine E. ;
Swift, Peter G. F. ;
Skinner, Chas T. ;
Aanstoot, Henk J. ;
Aman, Jan ;
Cameron, Fergus ;
Martul, Pedro ;
Chiarelli, Francesco ;
Daneman, Dennis ;
Danne, Thomas ;
Dorchy, Harry ;
Hoey, Hilary ;
Kaprio, Eero A. ;
Kaufman, Francine ;
Kocova, Mirjana ;
Mortensen, Henrik B. ;
Njolstad, Pal R. ;
Phillip, Moshe ;
Robertson, Kenneth J. ;
Schoenle, Eugen J. ;
Urakami, Tatsuhiko ;
Vanelli, Maurizio .
DIABETES CARE, 2007, 30 (09) :2245-2250
[10]   A Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes [J].
El-Khatib, Firas H. ;
Russell, Steven J. ;
Nathan, David M. ;
Sutherlin, Robert G. ;
Damiano, Edward R. .
SCIENCE TRANSLATIONAL MEDICINE, 2010, 2 (27) :27ra27