Systematic distortions in magnetic position digitizers

被引:93
作者
Birkfellner, W
Watzinger, F
Wanschitz, F
Enislidis, G
Kollmann, C
Rafolt, D
Nowotny, R
机构
[1] Univ Vienna, Gen Hosp, Dept Biomed Engn & Phys, Vienna, Austria
[2] Univ Vienna, Clin Oral & Maxillofacial Surg, Sch Med, Vienna, Austria
关键词
magnetic tracking systems; accuracy; frameless stereotaxy; computer aided surgery; 3D ultrasound;
D O I
10.1118/1.598425
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Medical devices equipped with position sensors enable applications like image guided surgical interventions, reconstruction of three-dimensional 3D ultrasound (US) images, and virtual or augmented reality systems. The acquisition of three-dimensional position data in real time is one of the key technologies in this field. The systematic distortions induced by various metals, surgical tools, and US scan probes in different commercial electromagnetic tracking systems were assessed in the presented work. A precise nonmetallic six degree-of-freedom measurement rack was built that allowed a quantitative comparison of different electromagnetic trackers. Also, their performance in the presence of large metallic structures was quantified in a phantom study on an acrylic skull model in an operating room (OR). The trackers used were alternating current (ac) and direct current (dc) based systems. The ac trackers were, on average, distorted by 0.7 mm and 0.5 degrees by metallic objects positioned at a distance greater than 120 mm between the geometrical center of the sample and the sensor. In the OR environment, the ac system exhibits mean errors of 3.2 +/- 2.4 mm and 2.9 degrees +/- 1.9 degrees. The de trackers are more sensitive to distortions caused by ferromagnetic materials (averaged value: 1.6 mm and 0.5 degrees beyond a distance of 120 mm). The de tracker shows no distortions from other conductive materials but was less accurate in the OR environment (typical error: 6.4 +/- 2.5 mm and 4.9 degrees +/- 2.0 degrees). At distances smaller than approximately 100 mm between sample and sensor error increases quickly. It is also apparent from our measurements that the influence of US scan probes is governed by their shielding material. The results show that surgical instruments not containing conductive material are to be preferred when using an ac tracker. Nonferromagnetic instruments should be used with de trackers. Static distortions caused by the OR environment have to be compensated by precise calibration methods. (C) 1998 American Association of Physicists in Medicine [S0094-2405(98)01611-3].
引用
收藏
页码:2242 / 2248
页数:7
相关论文
共 30 条
[1]  
BRYSON S, 1992, P SOC PHOTO-OPT INS, V1669, P244, DOI 10.1117/12.60417
[2]  
Bzostek A, 1997, LECT NOTES COMPUT SC, V1205, P299, DOI 10.1007/BFb0029249
[3]   COMPUTER-ASSISTED MEDICAL INTERVENTIONS - PASSIVE AND SEMIACTIVE AIDS [J].
CINQUIN, P ;
BAINVILLE, E ;
BARBE, C ;
BITTAR, E ;
BOUCHARD, V ;
BRICAULT, I ;
CHAMPLEBOUX, G ;
CHENIN, M ;
CHEVALIER, L ;
DELNONDEDIEU, Y ;
DESBAT, L ;
DESSENNE, V ;
HAMADEH, A ;
HENRY, D ;
LAIEB, N ;
LAVALLEE, S ;
LEFEBVRE, JM ;
LEITNER, F ;
MENGUY, Y ;
PADIEU, F ;
PERIA, O ;
POYET, A ;
PROMAYON, M ;
ROUAULT, S ;
SAUTOT, P ;
TROCCAZ, J ;
VASSAL, P .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1995, 14 (03) :254-263
[4]   3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION [J].
DETMER, PR ;
BASHEIN, G ;
HODGES, T ;
BEACH, KW ;
FILER, EP ;
BURNS, DH ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) :923-936
[5]   Computed intraoperative navigation guidance - a preliminary report on a new technique [J].
Enislidis, G ;
Wagner, A ;
Ploder, O ;
Ewers, R .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 1997, 35 (04) :271-274
[6]  
F Colchester at al A.C., 1994, Proceedings of Visualization in Biomedical Computing, P541
[7]   Image-guided endoscopic surgery: Results of accuracy and performance in a multicenter clinical study using an electromagnetic tracking system [J].
Fried, MP ;
Kleefield, J ;
Gopal, H ;
Reardon, E ;
Ho, BT ;
Kuhn, FA .
LARYNGOSCOPE, 1997, 107 (05) :594-601
[8]  
GHAZISAEDY M, 1995, VIRTUAL REALITY ANNUAL INTERNATIONAL SYMPOSIUM '95, PROCEEDINGS, P179
[9]   Intragastric distribution and gastric emptying assessed by three-dimensional ultrasonography [J].
Gilja, OH ;
Detmer, PR ;
Jong, JM ;
Leotta, DF ;
Li, XN ;
Beach, KW ;
Martin, R ;
Strandness, DE .
GASTROENTEROLOGY, 1997, 113 (01) :38-49
[10]  
GOERSS SJ, 1994, STEREOT FUNCT NEUROS, V63, P89, DOI 10.1159/000100299