MTF evaluation of a phosphor-coated CCD for x-ray imaging

被引:34
作者
Gambaccini, M [1 ]
Taibi, A [1 ]
DelGuerra, A [1 ]
Marziani, M [1 ]
Tuffanelli, A [1 ]
机构
[1] IST NAZL FIS NUCL,I-44100 FERRARA,ITALY
关键词
D O I
10.1088/0031-9155/41/12/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel digital x-ray detector has been assembled. The imaging system is based on a phosphor-coated charge-coupled device (CCD) obtained by direct deposition of a gadolinium oxysulphide scintillator onto the detector surface. The modulation transfer function has been measured along the two directions of the digital coordinates with the narrow slit technique. A resolution limit of about 20 line pairs per mm has been obtained for both directions. The high spatial resolution currently demanded in mammography can be achieved with this imaging system.