Corrosion behavior of nitinol wires in body fluid environments

被引:62
作者
Carroll, WM [1 ]
Kelly, MJ [1 ]
机构
[1] Natl Univ Ireland, Dept Chem, Galway, Ireland
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A | 2003年 / 67A卷 / 04期
关键词
nitinol; corrosion; body fluids; surface oxide; galvanic corrosion;
D O I
10.1002/jbm.a.10099
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, breakdown potentials were measured for unpolished and mechanically polished nitinol wires in simulated body fluids. These wires are similar to those used in the manufacture of stents. Considerable scatter was observed in the results indicating a variable surface state. After appropriate heat treatments, the measured breakdown values were lower but more reproducible for the mechanically polished samples. Significantly higher breakdown potentials were observed for cross-section wire samples. Some wires were tested in human blood and the breakdown values were higher than in Ringer and 0.9% NaCl solutions. Energy dispersive X-ray analysis of the surface layers indicated that oxide thickening occurred after heat treatments. Dynamic secondary ion mass spectroscopy also revealed thickened surface oxides on the wires. The oxide was predominantly made up of TiO2 with a very thin layer of NiO at the outer surface. Galvanic corrosion tests were performed on nitinol wires coupled with gold, elgiloy/phynox, and stainless steel. Nitinol was found to be anodic in all cases yet the currents measured were small. In tests in which nitinol-gold couples were immersed in 0.9% NaCl for periods up to 12 months, only very small amounts of nickel (in the part per billion range) were released into solution and scanning electron microscopy examination revealed no corrosion. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:1123 / 1130
页数:8
相关论文
共 22 条
[1]   INFLUENCE OF CATHODE-TO-ANODE AREA RATIO AND SEPARATION DISTANCE ON GALVANIC CORROSION CURRENTS OF STEEL IN CONCRETE CONTAINING CHLORIDES [J].
ARYA, C ;
VASSIE, PRW .
CEMENT AND CONCRETE RESEARCH, 1995, 25 (05) :989-998
[2]   BIODEGRADATION OF ORTHODONTIC APPLIANCES .1. BIODEGRADATION OF NICKEL AND CHROMIUM INVITRO [J].
BARRETT, RD ;
BISHARA, SE ;
QUINN, JK .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (01) :8-14
[3]   EFFECT OF LOW-TEMPERATURE PHASE CHANGES ON MECHANICAL PROPERTIES OF ALLOYS NEAR COMPOSITION TINI [J].
BUEHLER, WJ ;
WILEY, RC ;
GILFRICH, JV .
JOURNAL OF APPLIED PHYSICS, 1963, 34 (05) :1475-&
[4]   A CREVICE-FREE ELECTRODE ASSEMBLY FOR THE DETERMINATION OF REPRODUCIBLE BREAKDOWN POTENTIALS FOR STAINLESS-STEELS IN HALIDE ENVIRONMENTS [J].
CARROLL, WM ;
LYNSKEY, EE .
CORROSION SCIENCE, 1994, 36 (09) :1667-1678
[5]   OXIDATION OF AN NITI ALLOY [J].
CHAN, CM ;
TRIGWELL, S ;
DUERIG, T .
SURFACE AND INTERFACE ANALYSIS, 1990, 15 (06) :349-354
[6]  
FONTANA MG, 1967, CORROSION ENG, P28
[7]   Interactions of orthopaedic metals with an immortalized rat osteoblast cell line [J].
McKay, GC ;
Macnair, R ;
MacDonald, C ;
Grant, MH .
BIOMATERIALS, 1996, 17 (13) :1339-1344
[8]  
MonteroOcampo C, 1996, J BIOMED MATER RES, V32, P583, DOI 10.1002/(SICI)1097-4636(199612)32:4<583::AID-JBM11>3.0.CO
[9]  
2-F
[10]  
OSHIDA Y, 1991, CORROSION ENG, V40, P1009