摘要
Near infrared (NIR) spectroscopy is used for the rapid determination of decyl glucopyranoside (10G1) and alkyl polyglycoside (AG) mixtures having different alkyl chain length and the number of glucose unit in aqueous solution. NIR spectroscopy is a much simpler spectroscopic analysis method compared to three analytical methods for mixture of AGs such as thin layer chromatography (TLC), high performance liquid chromatography (HPLC), and photometry method. NIR spectra of AGs between 0.030 and 0.540 mg/ml in aqueous solutions were utilized to develop a calibration model. Both raw spectra and the second derivatives of AGs were tested for the best fit. The best calibration was built with second derivative spectra by using multiple linear regression (MLR). The standard error of calibration (SEC) and the standard error of prediction (SEP) were used for the evaluation of the model. The best calibration provides an SEP of 0.052 and 0.061 mg/ml for the prediction set of 10G1 and AG mixture, respectively. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:73 / 78
页数:6
相关论文
共 17 条
[1]   DETERMINATION OF PHYSIOLOGICAL LEVELS OF GLUCOSE IN AN AQUEOUS MATRIX WITH DIGITALLY FILTERED FOURIER-TRANSFORM NEAR-INFRARED SPECTRA [J].
ARNOLD, MA ;
SMALL, GW .
ANALYTICAL CHEMISTRY, 1990, 62 (14) :1457-1464
[2]  
Busch P, 1993, Tenside Surfactants Deterg, V30, P116
[3]  
Buschmann N, 1995, TENSIDE SURFACT DET, V32, P336
[4]  
BUSCHMANN N, 1996, AGRO FOOD IND HI TEC, V12, P6
[5]   Simultaneous measurements of glucose, glutamine, ammonia, lactate, and glutamate in aqueous solutions by near-infrared spectroscopy [J].
Chung, H ;
Arnold, MA ;
Rhiel, M ;
Murhammer, DW .
APPLIED SPECTROSCOPY, 1996, 50 (02) :270-276
[6]   In situ near-IR cure monitoring of a model epoxy matrix composite [J].
Cossins, S ;
Connell, M ;
Cross, B ;
Winter, R ;
Kellar, J .
APPLIED SPECTROSCOPY, 1996, 50 (07) :900-905
[7]  
Fischer E., 1893, CHEM BER, V26, P2400
[8]   INCREASING SENSITIVITY OF ANTHRONE METHOD FOR CARBOHYDRATE [J].
JERMYN, MA .
ANALYTICAL BIOCHEMISTRY, 1975, 68 (01) :332-335
[9]  
Klaffke HS, 1998, TENSIDE SURFACT DET, V35, P108
[10]  
Klaffke S, 1999, TENSIDE SURFACT DET, V36, P178