Multi-analyte high performance liquid chromatography coupled to high resolution tandem mass spectrometry method for control of pesticide residues, mycotoxins, and pyrrolizidine alkaloids

被引:110
作者
Dzuman, Zbynek [1 ]
Zachariasova, Milena [1 ]
Veprikova, Zdenka [1 ]
Godula, Michal [2 ]
Hajslova, Jana [1 ]
机构
[1] Univ Chem & Technol, Prague 16628 6, Czech Republic
[2] Thermo Sci, Prague 10000 10, Czech Republic
关键词
Pesticide residues; Mycotoxins; Quick easy cheap effective rugged safe; Orbitrap; High performance liquid chromatography; High resolution tandem mass spectrometry; TARGETED ANALYSIS; RECENT TRENDS; PLANT TOXINS; FOOD; EXTRACTION; FRUITS; FEED; METABOLITES; VALIDATION; CEREALS;
D O I
10.1016/j.aca.2015.01.021
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
A new reliable and highly sensitive method based on high performance liquid chromatographic (HPLC) separation and high resolution tandem mass spectrometric detection (HRMS/MS) has been developed and validated for determination of 323 pesticide residues, 55 mycotoxins, and 11 plant toxins represented by pyrrolizidine alkaloids. The method was validated for three matrices, leek, wheat, and tea differing in nature/amount of co-extracts that may cause various matrix effects. For target analytes isolation, optimized QuEChERS-based (quick, easy, cheap, effective, rugged, and safe) extraction procedure was employed. Spectral HRMS/MS library has been established providing an entire spectrum of fragment ions for each analyte, which allows unbiased identification and confirmation of target compounds. The limits of quantification (LOQs) of target analytes were below 10 mu g kg(-1) for 82%, 81%, and 61% for matrices leek, wheat, and tea, respectively. Recoveries were in the acceptable range (70-120%) according to SANCO/12571/2013 for most of target analytes, except for highly polar 'masked' mycotoxin deoxynivalenol-3-glucoside with recoveries 35%, 47%, and 42% for matrices leek, wheat, and tea, respectively. The linearities of calibration curves expressed as coefficients of determination were in the range of 0.9661-1.000, and repeatabilities expressed as relative standard deviations (RSDs) at LOQs lied in the range of 0.25-13.51%. The trueness of the method was verified using several certified reference materials (CRMs) and proficiency test samples. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:29 / 40
页数:12
相关论文
共 32 条
[1]
Al-Saleh Iman A., 1994, Journal of Environmental Pathology Toxicology and Oncology, V13, P151
[2]
Anastassiades M, 2003, J AOAC INT, V86, P412
[3]
[4]
[Anonymous], SANCO125712013 EUR C
[5]
Mycotoxins [J].
Bennett, JW ;
Klich, M .
CLINICAL MICROBIOLOGY REVIEWS, 2003, 16 (03) :497-+
[6]
UNE-EN ISO/IEC 17025:2005-accredited method for the determination of pesticide residues in fruit and vegetable samples by LC-MS/MS [J].
Camino-Sanchez, F. J. ;
Zafra-Gomez, A. ;
Oliver-Rodriguez, B. ;
Ballesteros, O. ;
Navalon, A. ;
Crovetto, G. ;
Vilchez, J. L. .
FOOD ADDITIVES AND CONTAMINANTS PART A-CHEMISTRY ANALYSIS CONTROL EXPOSURE & RISK ASSESSMENT, 2010, 27 (11) :1532-1544
[7]
Multi-residue determination of 130 multiclass pesticides in fruits and vegetables by gas chromatography coupled to triple quadrupole tandem mass spectrometry [J].
Cervera, M. I. ;
Medina, C. ;
Portoles, T. ;
Pitarch, E. ;
Beltran, J. ;
Serrahima, E. ;
Pineda, L. ;
Munoz, G. ;
Centrich, F. ;
Hernandez, F. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 397 (07) :2873-2891
[8]
A rugged high-throughput analytical approach for the determination and quantification of multiple mycotoxins in complex feed [J].
Dzuman, Zbynek ;
Zachariasova, Milena ;
Lacina, Ondrej ;
Veprikova, Zdenka ;
Slavikova, Petra ;
Hajslova, Jana .
TALANTA, 2014, 121 :263-272
[9]
Henderson W., 2005, MASS SPECTROMETRY IN
[10]
Recent developments in liquid chromatography-mass spectrometry and related techniques [J].
Holcapek, Michal ;
Jirasko, Robert ;
Lisa, Miroslav .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1259 :3-15