Determination of erythrocyte deformability and its correlation to cellular ATP release using microbore tubing with diameters that approximate resistance vessels in vivo

被引:38
作者
Fischer, DJ [1 ]
Torrence, NJ [1 ]
Sprung, RJ [1 ]
Spence, DM [1 ]
机构
[1] St Louis Univ, Dept Chem, St Louis, MO 63103 USA
关键词
D O I
10.1039/b308225n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel method is described for measuring the deformability of red blood cells (RBCs) in tubing whose diameters approximate forces encountered in vivo. Here, RBCs from rabbits are loaded into a 50 cm section of 75 mm id microbore tubing and connected to a syringe pump. This section of tubing is then connected to a 15 cm section of 25 mum id tubing. As buffer is pumped through the flow system, the RBCs are evacuated from both sections of tubing. However, the inability of the RBCs to move freely through the 25 mum id section of tubing results in a buildup of cells at the inlet of this portion of tubing. The continued force output by the syringe pump results in a deformation of the RBCs until all of the cells are eventually evacuated from the flow system. It was found that a measurement of the time required to reach half of the maximum pressure (1/2P(max)) may be used as an indicator of the RBC deformability. For a given sample, a simple buffer results in less time to reach 1/2P(max) (6.9 +/- 0.2 s) than deformable RBCs (21.6 +/- 0.8 s). To verify that the increased amount of time to reach 1/2P(max) is indeed due to the RBCs, various hematocrits of an RBC sample were investigated and, as expected, it was found that a 12% RBC hematocrit had a higher 1/2P(max) value (26.0 s +/- 2.2 s) when compared to a 7% hematocrit (19.1 +/- 0.3 s). In addition, RBCs chemically stiffened with glutaraldehyde were shown to be 25% less deformable than normal RBCs. Finally, a study was performed to examine the relationship between RBC deformability and ATP release and it was found that ATP release increased as a function of RBC deformability. This method greatly simplifies deformability measurements, employing only a syringe pump and microbore tubing, and may lead to a more complete understanding of the physiological significance of erythrocyte deformability.
引用
收藏
页码:1163 / 1168
页数:6
相关论文
共 23 条
[1]  
[Anonymous], 1998, Experimental and Clinical Cardiology
[2]   MECHANISMS OF CORONARY VASODILATATION PRODUCED BY ATP IN GUINEA-PIG ISOLATED PERFUSED HEART [J].
BROWN, IP ;
THOMPSON, CI ;
BELLONI, FL .
BRITISH JOURNAL OF PHARMACOLOGY, 1992, 105 (01) :211-215
[3]  
DORMANDY J, 1985, CLIN HEMORHEOL, V5, P975
[4]   CAPILLARY RESISTANCE TO FLOW OF HARDENED (DIAMIDE TREATED) RED-BLOOD-CELLS (RBC) [J].
DRIESSEN, GK ;
SCHEIDTBLEICHERT, H ;
SOBOTA, A ;
INHOFFEN, W ;
HEIDTMANN, H ;
HAEST, CWM ;
KAMP, D ;
SCHMIDSCHONBEIN, H .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1982, 392 (03) :261-267
[5]   Chemiluminescence detection of ATP release from red blood cells upon passage through microbore tubing [J].
Edwards, J ;
Sprung, R ;
Sprague, R ;
Spence, D .
ANALYST, 2001, 126 (08) :1257-1260
[6]   THE OBLIGATORY ROLE OF ENDOTHELIAL-CELLS IN THE RELAXATION OF ARTERIAL SMOOTH-MUSCLE BY ACETYLCHOLINE [J].
FURCHGOTT, RF ;
ZAWADZKI, JV .
NATURE, 1980, 288 (5789) :373-376
[7]  
GOLDSMITH HL, 1968, J GEN PHYSIOL, V52, pS5
[8]  
GOLDSMITH HL, 1971, THEORETICAL CLINICAL, P47
[9]  
HARDEMAN MR, 1994, CLIN HEMORHEOL, V14, P605
[10]  
Hardeman MR, 1999, CLIN HEMORHEOL MICRO, V21, P277