Portable gamma spectrometry with cerium-doped lanthanum bromide scintillators: Suitability assessments for luminescence and electron spin resonance dating applications

被引:72
作者
Arnold, L. J. [1 ]
Duval, M. [1 ]
Falgueres, C. [2 ]
Bahain, J. -J. [2 ]
Demuro, M. [1 ]
机构
[1] CENIEH, Ctr Nacl Invest Sobre Evoluc Humana, Burgos 09002, Spain
[2] Museum Natl Hist Nat, Dept Prehist, CNRS, UMR7194, F-75013 Paris, France
关键词
Lanthanum bromide; Sodium iodide; Scintillator detectors; Portable gamma ray spectrometry; Luminescence dating; Electron spin resonance dating; NAI(TL) SCINTILLATORS; LABR3-CE; CRYSTALS; RAY;
D O I
10.1016/j.radmeas.2011.09.001
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
070305 [高分子化学与物理]; 080810 [电力电子与电能变换];
摘要
Cerium-doped lanthanum bromide (LaBr3:Ce) crystals offer a range of improved scintillation properties over traditional NaI:Tl crystals for in situ gamma spectrometry. At present, however, it remains unclear whether the internal radioactivity of LaBr3:Ce detectors compromises their suitability for low-level activity radioisotope measurements of natural sedimentary deposits, such as those required in luminescence and electron spin resonance (ESR) dating. In this study we investigate the suitability of a commercial LaBr3:Ce detector for measuring individual concentrations of K-40, U-238 and Th-232 using predefined 'energy windows' from gamma ray spectra. Performance tests have been undertaken using reference materials with well-constrained radioisotope concentrations (the Oxford calibration blocks) and compared with results obtained for a NaI:Tl detector of the same geometry. These tests reveal that the LaBr3:Ce detector has a non-negligible intrinsic activity that needs to be accurately quantified prior to measuring any gamma ray spectra in the field. Compared to the NaI:Tl detector, the energy resolution of the LaBr3:Ce detector is improved by a factor of two, or more, for the main indicator isotope photopeaks in the K-40, U-238 and Th-232 decay series. Signal-to-noise ratios for the LaBr3:Ce detector show a 25-35% improvement over those of the NaI:Tl detector. In addition, the LaBr3:Ce detector is characterised by suitable energy linearity over the full spectral range of interest for the K-40, U-238 and Th-232 decay series. Replicate gamma ray measurements made with the LaBr3:Ce and NaI:Tl detectors for 20 natural sedimentary samples from the Lower Tejo River basin, Portugal, and the Duero River basin, Spain, yield consistent radioisotope concentrations and gamma dose rate estimates. These results are encouraging and suggest that LaBr3:Ce detectors can provide suitable estimates of individual radioisotope concentrations in low-level activity (0.5-1.5 Gy/ka) environments, providing that their intrinsic activity is adequately measured and subtracted from field spectra. Our comparison also reveals that subtraction of the intrinsic activity from LaBr3:Ce spectra produces a significant reduction in the precision with which radionuclide concentrations can be determined using the 'energy windows' approach. This shortcoming necessitates longer counting times in natural sedimentary environments and overshadows the practical advantages that LaBr3:Ce detectors might otherwise offer for luminescence and ESR dating applications. (c) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:6 / 18
页数:13
相关论文
共 23 条
[1]
Adamiec G., 1998, ANCIENT TL, V16, P37, DOI DOI 10.1016/S0277-3791(03)00021-0
[2]
Aitken M.J., 1985, Thermoluminescence Dating
[3]
[Anonymous], THESIS U OXFORD
[4]
Benito Calvo A., 2009, EARTH SURF PROCESSES, V34, P1433
[5]
Benito Calvo A., 2004, THESIS U COMPLUTENSE
[6]
Ekstrom L.P., 1999, WWW Table of Radioactive Isotopes
[7]
Hubbell J.H., 2019, XRAY MASS ATTENUATIO
[8]
Influence of the background approximation methods on the analysis of γ-ray spectra [J].
Luca, A ;
Morel, J .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (2-4) :233-237
[9]
Geoarchaeology of Pleistocene open-air sites in the Vila Nova da Barquinha-Santa Cita area (Lower Tejo River basin, central Portugal) [J].
Martins, Antonio A. ;
Cunha, Pedro P. ;
Rosina, Pierluigi ;
Osterbeek, Luiz ;
Cura, Sara ;
Grimaldi, Stefano ;
Gomes, Jose ;
Buylaert, Jan-Pieter ;
Murray, Andrew S. ;
Matos, Joao .
PROCEEDINGS OF THE GEOLOGISTS ASSOCIATION, 2010, 121 (02) :128-140
[10]
Performance of large lanthanum bromide scintillators [J].
Menge, Peter R. ;
Gautier, G. ;
Iltis, A. ;
Rozsa, C. ;
Solovyev, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01) :6-10