Designing passive MRI-safe implantable conducting leads with electrodes

被引:68
作者
Bottomley, Paul A. [1 ]
Kumar, Ananda [1 ]
Edelstein, William A. [1 ]
Allen, Justin M. [1 ]
Karmarkar, Parag V. [1 ]
机构
[1] SurgiVision Inc, Baltimore, MD 21218 USA
关键词
MRI safety; implanted leads; RF heating; MRI-safe leads; DEEP BRAIN-STIMULATION; CARDIOVERTER-DEFIBRILLATORS; CARDIAC-PACEMAKERS; ABSORPTION RATE; HEATING TESTS; RF SAFETY; WIRES; PERFORMANCE; CATHETERS; DEVICES;
D O I
10.1118/1.3439590
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: The presence of implanted electronic devices with conducting leads and electrodes are contraindicated for magnetic resonance imaging (MRI), denying many patients its potential benefits. The prime concern is MRI's radio frequency (RF) fields, which can cause elevated local specific absorption rates (SARs) and potential heat injury. The purpose of this article is to develop and compare a range of passive implantable "MRI-safe" lead designs. Methods: Conducting leads incorporating different lengths (3-75cm), insulation thicknesses (0-105 mu m), resistances (100-3000 Omega), coiled conductors (inner diameter <= 1.2 mm), high-impedance (135-2700 Omega) RF traps, and single-coiled and triple-coiled coaxial-wound "billabong" leads with reversed coil sections that oppose and reduce the induced current, are investigated both experimentally using local temperature measurements, and by numerical full-wave electromagnetic field analysis of the local SAR, in three different-sized bioanalogous model saline-gel phantoms at 1.5 T MRI and 4 W/kg exposure. Results: In all designs, the maximum computed 1 g average SAR and experimental temperature rise occur at the bare electrodes. Electrode heating increases with lead insulation thickness and peaks for uncoiled leads 25-50 cm long. A reasonable match between computed SAR and the point SAR estimated from thermal sensors obtained by approximating the computation volume to that of the thermal probes. Factors that maximize the impedance of leads with resistive, coiled, RF trap and billabong elements can effectively limit heating below 1-2 degrees C, but folded lead configurations can be a concern. The RF trap and billabong designs can both support multiple conductors and electrodes, with billabong prototype leads also heating <1 degrees C when tested for 3 T MRI. Conclusions: Lead insulation and length strongly affect implanted lead safety to RF exposure during MRI. Lead designs employing impedance and reversed winding sections offer hope for the development of passive, MRI-safe, implantable conducting leads for future human use. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3439590]
引用
收藏
页码:3828 / 3843
页数:16
相关论文
共 60 条
[1]
Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]
*ASTM INT, 2009, WK21326 ASTM INT
[3]
High resolution intravascular MRI and MRS by using a catheter receiver coil [J].
Atalar, E ;
Bottomley, PA ;
Ocali, O ;
Correia, LCL ;
Kelemen, MD ;
Lima, JAC ;
Zerhouni, EA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (04) :596-605
[4]
ATALAR E, 1999, P INT SOC MAG RESON, V7, P1006
[5]
Variability in RF-induced heating of a deep brain stimulation implant across MR systems [J].
Baker, Kenneth B. ;
Tkach, Jean A. ;
Phillips, Micheal D. ;
Rezai, Ali R. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (06) :1236-1242
[6]
Bottomley P. A., 2009, P INT SOC MAG RESON, V17, P4805
[7]
HOMOGENEOUS TISSUE MODEL ESTIMATES OF RF POWER DEPOSITION IN HUMAN NMR-STUDIES - LOCAL ELEVATIONS PREDICTED IN SURFACE COIL DECOUPLING [J].
BOTTOMLEY, PA ;
ROEMER, PB .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES-SERIES, 1992, 649 :144-159
[8]
RF MAGNETIC-FIELD PENETRATION, PHASE-SHIFT AND POWER DISSIPATION IN BIOLOGICAL TISSUE - IMPLICATIONS FOR NMR IMAGING [J].
BOTTOMLEY, PA ;
ANDREW, ER .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (04) :630-643
[9]
Bottomley PA., 2001, SAFETY INTERNAL MRI, P85
[10]
In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: Role of implant geometry [J].
Calcagnini, Giovanni ;
Triventi, Michele ;
Censi, Federica ;
Mattei, Eugenio ;
Bartolini, Pietro ;
Kainz, Wolfgang ;
Bassen, Howard I. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (04) :879-886