High-speed, high-sensitivity aerosol fluorescence spectrum detection using a 32-anode photomultiplier tube detector

被引:42
作者
Pan, YL [1 ]
Cobler, P
Rhodes, S
Potter, A
Chou, T
Holler, S
Chang, RK
Pinnick, RG
Wolf, JP
机构
[1] New Mexico State Univ, Phys Sci Lab, Las Cruces, NM 88003 USA
[2] 20 New England Business Ctr, Vtech Engn Corp, Andover, MA 01810 USA
[3] Yale Univ, Dept Appl Phys, New Haven, CT 06520 USA
[4] Yale Univ, Ctr Laser Diagnost, New Haven, CT 06520 USA
[5] USA, Res Lab, Computat & Informat Sci Directorate, Adelphi, MD 20783 USA
[6] Univ Lyon 1, LASIM, CNRS, UMR 5579, F-69622 Villeurbanne, France
关键词
D O I
10.1063/1.1344179
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We employ a 32-anode photomultiplier tube (PMT) in a fluorescence detection system and demonstrate its ability to record broad fluorescence spectra at frame rates in excess of 1.4x10(3) Hz, which is 56x faster than the frame rate of an intensified charge coupled device detector. The multi-anode PMT has single-photon detectable sensitivity. A new data acquisition and processing system for the multi-anode PMT, together with the system-controlling software, has been developed. The performance characteristics of the fluorescence detection system, including the data rate capability, dynamic range, signal-to-noise ratio, and crosstalk among the different anodes, are reported. The 32-anode PMT and acquisition system are suitable for a real-time, field-portable, multichannel optical analyzer. (C) 2001 American Institute of Physics.
引用
收藏
页码:1831 / 1836
页数:6
相关论文
共 15 条
[1]  
BOTTIGER JR, 1998, J AEROSOL SCI, V29, pS965
[2]   Conditional-firing aerosol-fluorescence spectrum analyzer for individual airborne particles with pulsed 266-nm laser excitation [J].
Chen, G ;
Nachman, P ;
Pinnick, RG ;
Hill, SC ;
Chang, RK .
OPTICS LETTERS, 1996, 21 (16) :1307-1309
[3]   Detection of bioaerosols using multiwavelength UV fluorescence spectroscopy [J].
Cheng, YS ;
Barr, EB ;
Fan, BJ ;
Hargis, PJ ;
Rader, DJ ;
O'Hern, TJ ;
Torczynski, JR ;
Tisone, GC ;
Preppernau, BL ;
Young, SA ;
Radloff, RJ .
AEROSOL SCIENCE AND TECHNOLOGY, 1999, 30 (02) :186-201
[4]  
Eversole JD, 1999, FIELD ANAL CHEM TECH, V3, P249, DOI 10.1002/(SICI)1520-6521(1999)3:4/5<249::AID-FACT4>3.0.CO
[5]  
2-O
[6]   Design of an instrument for real-time detection of bioaerosols using simultaneous measurement of particle aerodynamic size and intrinsic fluorescence [J].
Hairston, PP ;
Ho, J ;
Quant, FR .
JOURNAL OF AEROSOL SCIENCE, 1997, 28 (03) :471-482
[7]  
*HAM PHOT KK ED CO, PHOT TUB PRINC APPL
[8]  
Hill SC, 1999, FIELD ANAL CHEM TECH, V3, P221, DOI 10.1002/(SICI)1520-6521(1999)3:4/5<221::AID-FACT2>3.3.CO
[9]  
2-Z
[10]   Simultaneous light scattering and intrinsic fluorescence measurement for the classification of airborne particles [J].
Kaye, PH ;
Barton, JE ;
Hirst, E ;
Clark, JM .
APPLIED OPTICS, 2000, 39 (21) :3738-3745