An EGSnrc investigation of cavity theory for ion chambers measuring air kerma

被引:22
作者
Buckley, LA [1 ]
Kawrakow, I
Rogers, DWO
机构
[1] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
关键词
dosimetry; Monte Carlo; ionization chamber;
D O I
10.1118/1.1573891
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The EGSnrc system is used to compare the response of an aluminum-walled thimble chamber to that of a graphite-walled thimble chamber for a Co-60 beam. When compared to previous experimental results, the EGSnrc values of the ratios of chamber response differ by as much as 0.7% from the experiment. However, it is shown that this difference can be more than accounted for by switching from using the graphite mean excitation energy of 78 eV used in dosimetry protocols to the value of 86.8 eV suggested by more recent stopping-power experiments. This suggests that the uncertainty analysis of Monte Carlo results must be done more carefully, by taking into account uncertainties in the underlying basic data such as the electron and photon cross sections. In comparison to Spencer-Attix cavity theory for a thick-walled ion chamber, the Monte Carlo calculated values of the chamber response differ from the expected ones by 0.15% and 0.01% for the graphite and aluminum chambers, respectively, which are comparable to previously reported values for the Spencer-Attix correction factors. EGSnrc is also used to investigate the effect on the chamber response of thin dag layers on the inside of the aluminum wall. There is good agreement between the calculated and measured changes in chamber response versus the thickness of the dag. The results are compared to the predictions of the Almond-Svensson extension of cavity theory and show that the theory does not correctly predict the chamber response in the presence of thin dag layers. This finding is in agreement with previously reported experimental results. It is demonstrated that the values of a, the fraction of ionizations in the gas arising from electrons generated in the dag layer, used in the theory, are not the source of the disagreement.
引用
收藏
页码:1211 / 1218
页数:8
相关论文
共 28 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   IONIZATION-CHAMBER DOSIMETRY FOR PHOTON AND ELECTRON-BEAMS - THEORETICAL CONSIDERATIONS [J].
ALMOND, PR ;
SVENSSON, H .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1977, 16 (02) :177-186
[3]  
[Anonymous], 1983, MED PHYS, V10, P741
[4]  
[Anonymous], PIRS702 NAT RES COUN
[5]  
BICHSEL H, 1992, NUCL INSTRUM METH B, V66, P345
[6]   IMPLICATIONS OF NEW CORRECTION FACTORS ON PRIMARY AIR KERMA STANDARDS IN CO-60 BEAMS [J].
BIELAJEW, AF ;
ROGERS, DWO .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1283-1291
[7]   Monte Carlo study of correction factors for Spencer-Attix cavity theory at photon energies at or above 100 keV [J].
Borg, J ;
Kawrakow, I ;
Rogers, DWO ;
Seuntjens, JP .
MEDICAL PHYSICS, 2000, 27 (08) :1804-1813
[8]  
Hubbell J. H., 1995, 5632 NISTIR
[9]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[10]  
*IAEA, 1987, TECHN REP SER IAEA, V277