Suggestion of potential stent design parameters to reduce restenosis risk driven by foreshortening or dogboning due to non-uniform balloon-stent expansion

被引:51
作者
Lim, Dohyung [1 ]
Cho, Seung-Kwan [1 ]
Park, Won-Pil [1 ]
Kristensson, Anders [2 ]
Ko, Jai-Young [3 ]
Al-Hassani, S. T. S. [2 ]
Kim, Han-Sung [1 ]
机构
[1] Yonsei Univ, Res Inst Med Instruments & Rehabil, Dept Biomed Engn, Wonju 220710, Kangwondo, South Korea
[2] Univ Manchester, Dept Mech Aerosp & Manufacturing Engn, Manchester, Lancs, England
[3] MI Tech Co Ltd, Seoul, South Korea
关键词
stent; foreshortening; dogboning; transient non-uniform balloon expansion; finite element analysis;
D O I
10.1007/s10439-008-9504-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The foreshortening or dogboning of a stent that occurs due to transient non-uniform balloon-stent expansion can induce a vascular injury, resulting in restenosis of the coronary artery. However, previous studies rarely considered the effects of transient non-uniform balloon expansion on analysis of the mechanical properties and behaviors of stents during stent deployment, nor did they determine design parameters to minimize the restenosis risk driven by foreshortening or dogboning. The aim of the current study was, therefore, to suggest potential design parameters capable of reducing the possibility of restenosis risk driven by foreshortening or dogboning through a comparative study of seven commercial stents using finite element (FE) analyses of a realistic transient non-uniform balloon-stent expansion process. The results indicate that using stents composed of opened unit cells connected by bend-shaped link structures, in particular the MAC Plus stent, and controlling the geometrical and morphological features of the unit cell strut or the link structure at the distal ends of stent may prevent restenosis risk caused by foreshortening or dogboning. This study provides a first look at the realistic transient non-uniform balloon-stent expansion by investigating the mechanical properties, behaviors, and design parameters capable of reducing the possibility of restenosis risk induced by the foreshortening or the dogboning.
引用
收藏
页码:1118 / 1129
页数:12
相关论文
共 35 条
[1]
DYNAMIC UNIAXIAL AND BIAXIAL STRESS-STRAIN RELATIONSHIPS FOR AUSTENITIC STAINLESS-STEELS [J].
ALBERTINI, C ;
MONTAGNANI, M .
NUCLEAR ENGINEERING AND DESIGN, 1980, 57 (01) :107-123
[2]
Hemodynamics and wall mechanics of a compliance matching stent: In vitro and in vivo analysis [J].
Berry, JL ;
Manoach, E ;
Mekkaoui, C ;
Rolland, PH ;
Moore, JE ;
Rachev, A .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (01) :97-105
[3]
COLLAPSE OF A PALMAZ STENT IN THE SUBCLAVIAN VEIN [J].
BJARNASON, H ;
HUNTER, DW ;
CRAIN, MR ;
FERRAL, H ;
MILTZMILLER, SE ;
WEGRYN, SA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (05) :1123-1124
[4]
Carter AJ, 2000, CATHETER CARDIO INTE, V51, P112, DOI 10.1002/1522-726X(200009)51:1<112::AID-CCD26>3.0.CO
[5]
2-H
[6]
Effects of varying slotted tube (stent) geometry on its expansion behaviour using finite element method [J].
Chua, SND ;
MacDonald, BJ ;
Hashmi, MSJ .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2004, 155 :1764-1771
[7]
Finite element simulation of stent and balloon interaction [J].
Chua, SND ;
Mac Donald, BJ ;
Hashmi, MSJ .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2003, 143 :591-597
[8]
Finite-element simulation of stent expansion [J].
Chua, SND ;
Mac Donald, BJ ;
Hashmi, MSJ .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2002, 120 (1-3) :335-340
[9]
Management of restenosis after coronary intervention [J].
Dangas, G ;
Fuster, V .
AMERICAN HEART JOURNAL, 1996, 132 (02) :428-436
[10]
David ChuaS N., 2004, J MATER PROCESS TE 2, V155-156, P1772