A new assay design for clinical diagnostics based on alternative recognition elements

被引:20
作者
Albrecht, Christiane [1 ]
Fechner, Peter [1 ]
Honcharenko, Dmytro [2 ]
Baltzer, Lars [2 ]
Gauglitz, Guenter [1 ]
机构
[1] Univ Tubingen, Inst Phys & Theoret Chem, D-72076 Tubingen, Germany
[2] Uppsala Univ, Dept Biochem & Organ Chem, S-75123 Uppsala, Sweden
关键词
Immunoassays; (biological) Recognition elements; Kinetics; Clinical diagnostics; IMMUNOASSAY FORMATS; ANTIBODIES;
D O I
10.1016/j.bios.2010.03.022
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Herein, we present a new sandwich assay design containing a high affinity polypeptide scaffold as immobilized capture element and an antibody for detection. These polypeptide scaffolds provide a good affinity towards one antigen and can be linked to biosensor surfaces without affecting their binding capabilities. Furthermore, the small peptides are very stable, which allows for regenerating the surface several hundreds of times and thus for reuse of the biosensor. Moreover, these receptors can be synthesized with different affinities towards one antigen, which has been proven by characterizing them using a label-free detection method RIfS (reflectometric interference spectroscopy) for collecting kinetic data. Polypeptide scaffolds with different affinities have been chosen and characterized. Upon these results, sandwich-type assays have been set-up using a fluorescently labelled antibody as detection element. Thereby could be shown, that the working range of the assay can be shifted according to the affinity of the used capturing polypeptide scaffold. The scaffolds with a higher affinity towards the antigen can detect lower concentration, and in contrary, scaffolds with lower affinities can detect higher concentrations. In consequence, using this new sandwich-type assay, we avoid the complex procedure to immobilize antibodies in correct orientation, but simultaneously keep this well-known recognition element in the assay for detection. Furthermore, in addition to all the acknowledged properties of immunoassays, we add the possibility of tuning the working range of assays in distinct manner according to request. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:2302 / 2308
页数:7
相关论文
共 22 条
[1]   Two immunoassay formats for fully automated CRP detection in human serum [J].
Albrecht, Christiane ;
Kaeppel, Nina ;
Gauglitz, Guenter .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 391 (05) :1845-1852
[2]  
Atkins P.W., 2006, Physical Chemistry, V8
[3]   Polypeptide conjugate binders for protein recognition [J].
Baltzer, Lars .
CREATIVE CHEMICAL SENSOR SYSTEMS, 2007, 277 :89-106
[4]  
BARLEN B, 2009, PHYS STATUS SOLIDI A, P409
[5]   Optical multiple-analyte immunosensor for water pollution control [J].
Barzen, C ;
Brecht, A ;
Gauglitz, G .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (04) :289-295
[6]  
Ramírez NB, 2009, BRAZ J CHEM ENG, V26, P227, DOI 10.1590/S0104-66322009000200001
[7]   Antibodies in diagnostics - from immunoassays to protein chips [J].
Borrebaeck, CAK .
IMMUNOLOGY TODAY, 2000, 21 (08) :379-382
[8]  
BRECHT A, 1998, OPTICAL IMMUNOPROBE, P69
[9]  
DUDLEY RA, 1985, CLIN CHEM, V31, P1264
[10]  
EDDOWES MJ, 1987, BIOSENSORS, V3, P1