Reagent method sheets for analysis of common serum analytes often highlight the possibility of interference from lipaemia but the information given is often brief and may not be instrument-specific. Thus study:assesses the degree of interference from lipaemia in a range of common serum analytes on the Bayer Opera (with a serum blank) using a commercial, polymer, LipoClear, as a lipid-clearing agent, Serum samples (mean serum triglyceride 6.89 [range 0.58-28.4] mmol/L) are analysed for 14 common chemistry analytes and the results compared before and after treatment with LipoClear. Results showed no significant critical differences in analyte values before and, after treatment, except for an expected fall in total protein, phosphate, cholesterol and triglyceride concentrations. Most of the common analytes in use on the Bayer Opera are not subject to interference from lipaemia; however, we recommend that where method sheets indicate interference from lipaemia then this should be quantified for the analyte in question.