This study evaluated the wear property of four novel double-network (DN) hydrogels, which was composed of two kinds of hydrophilic polymers, using pin-on-flat wear testing. The gels involve PAMPS-PAAm get which consists of poly(2-acrylamide-2metyl-propane sulfonic acid) and polyacrylamide, PAMPS-PDAAAm gel which consists of poly(2-acrylamide-2-metyl-propane sulfonic acid) and poly(N,N'-dimetyl acrylamide), Cellulose/PDMAAm gel which consists of bacterial Cellulose and poly dimetylacrylamide, and Cellulose-Gelatin gel which consists of bacterial Cellulose and Gelatin. Ultra-high molecular weight polyethylene (UHMWPE) was used as a control of a clinically available material. Using a reciprocating apparatus, 10(6) Cycles of friction between a flat specimen and ceramic pin were repeated in water under a contact pressure of 0.1 MPa. To determine the depth and the roughness of the concave lesion created by wear, a confocal laser microscope was used. As a result, the maximum wear depth of the PAMPS-PDMAAm gel (3.20 mu m) was minimal in the five materials, while there was no significant difference compared to UHMWPE. There were significant differences between UHMWPE and one of the other three gels. The PAMPS-PAAm gel (9.50 mu m), the Cellulose PDMAAm gel (7.80 mu m), and the Cellulose-Gelatin gel (1302.40 mu m). This study demonstrated that the PAMPS-PDMAAm DN gel has an amazing wear property as a hydrogel that is comparable to the UHMWPE. In addition, the PAMPS-PAAm and Cellulose-PDMAAm DN gels are also resistant to wear to greater degrees than conventionally reported hydrogels. On the other hand, this study showed that the Cellulose-Gelatin DN gel was not resistant to wear. (c) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4468 / 4475
页数:8
相关论文
共 19 条
[1]  
*AM SOC TEST MAN, 2003, ANN BOOK ASTM STAND, P269
[2]  
DAISAKU K, IN PRESS ADV MAT, V16
[3]   Friction, wear, and lubrication of hydrogels as synthetic articular cartilage [J].
Freeman, ME ;
Furey, MJ ;
Love, BJ ;
Hampton, JM .
WEAR, 2000, 241 (02) :129-135
[4]   Double-network hydrogels with extremely high mechanical strength [J].
Gong, JP ;
Katsuyama, Y ;
Kurokawa, T ;
Osada, Y .
ADVANCED MATERIALS, 2003, 15 (14) :1155-+
[5]   Synthesis of hydrogels with extremely low surface friction [J].
Gong, JP ;
Kurokawa, T ;
Narita, T ;
Kagata, G ;
Osada, Y ;
Nishimura, G ;
Kinjo, M .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2001, 123 (23) :5582-5583
[6]   SYNTHESIS OF CELLULOSE BY ACETOBACTER-XYLINUM .2. PREPARATION OF FREEZE-DRIED CELLS CAPABLE OF POLYMERIZING GLUCOSE TO CELLULOSE [J].
HESTRIN, S ;
SCHRAMM, M .
BIOCHEMICAL JOURNAL, 1954, 58 (02) :345-352
[7]  
Kakugo A, 2002, ADV MATER, V14, P1124, DOI 10.1002/1521-4095(20020816)14:16<1124::AID-ADMA1124>3.0.CO
[8]  
2-M
[9]   Preliminary study of polyvinyl alcohol-hydrogel (PVA-H) artificial meniscus [J].
Kobayashi, M ;
Toguchida, J ;
Oka, M .
BIOMATERIALS, 2003, 24 (04) :639-647
[10]   FRICTION AND WEAR PROPERTIES OF POLYMER, METAL, AND CERAMIC PROSTHETIC JOINT MATERIALS EVALUATED ON A MULTICHANNEL SCREENING DEVICE [J].
MCKELLOP, H ;
CLARKE, I ;
MARKOLF, K ;
AMSTUTZ, H .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (05) :619-653