Detection of heavy metal ions at femtomolar levels using protein-based biosensors

被引:151
作者
Bontidean, I
Berggren, C
Johansson, G
Csöregi, E
Mattiasson, B
Lloyd, JA
Jakeman, KJ
Brown, NL
机构
[1] Univ Lund, Ctr Chem, Dept Biotechnol, S-22100 Lund, Sweden
[2] Univ Lund, Ctr Chem, Dept Analyt Chem, S-22100 Lund, Sweden
[3] Univ Birmingham, Sch Biol Sci, Birmingham B15 2TT, W Midlands, England
关键词
D O I
10.1021/ac9803636
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sensors based on proteins (GST-SmtA and MerR) with distinct binding sites for heavy metal ions were developed and characterized, A capacitive signal transducer was used to measure the conformational change following binding. The proteins were overexpressed in Escherichia coli, purified, and immobilized in different ways to a self-assembled thiol layer on a gold electrode placed as the working electrode in a potentiostatic arrangement in a flow analysis system. The selectivity and the sensitivity of the two protein-based biosensors were measured and compared for copper, cadmium, mercury, and zinc ions, The GST-SmtA electrodes displayed a broader selectivity (sensing all four heavy metal ions) compared with the MerR-based ones, which showed an accentuated selectivity for mercury ions. Metal ions could be detected with both electrode types down to femtomolar concentration. The upper measuring Limits, presumably due to near saturation of the proteins' binding sites, were around 10(-10) M. Control electrodes similarly constructed but based on bovine serum albumin or urease did not yield any signals. The electrodes could be regenerated with EDTA and used for more than 2 weeks with about 40% reduction in sensitivity.
引用
收藏
页码:4162 / 4169
页数:8
相关论文
共 35 条
[1]  
Anderson JL, 1996, ANAL CHEM, V68, pR379, DOI 10.1021/a1960015y
[2]  
ANDERSON JL, 1996, ANAL CHEM, V68, pR410
[3]   FORMATION OF MONOLAYER FILMS BY THE SPONTANEOUS ASSEMBLY OF ORGANIC THIOLS FROM SOLUTION ONTO GOLD [J].
BAIN, CD ;
TROUGHTON, EB ;
TAO, YT ;
EVALL, J ;
WHITESIDES, GM ;
NUZZO, RG .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1989, 111 (01) :321-335
[4]  
BATES RG, 1973, DETERMINATION PH
[5]   Capacitance measurements of antibody-antigen interactions in a flow system [J].
Berggren, C ;
Johansson, G .
ANALYTICAL CHEMISTRY, 1997, 69 (18) :3651-3657
[6]  
BERGGREN C, IN PRESS INSTRUM SCI
[7]  
BERGGREN C, 1998, IN PRESS BIOSENS BIO
[8]  
Beveridge TJ, 1997, ADV MICROB PHYSIOL, V38, P177
[9]  
BRADLEY CR, 1985, ANAL CHEM, V57, P1041
[10]  
Burlingame AL, 1996, ANAL CHEM, V68, pR599, DOI 10.1021/a1960021u