Determination of endogenous concentrations of the lanthanides in body fluids and tissues using electrothermal vaporization inductively coupled plasma mass spectrometry

被引:19
作者
Buseth, E
Wibetoe, G
Martinsen, I
机构
[1] Univ Oslo, Dept Chem, N-0315 Oslo, Norway
[2] Nycomed Imaging AS, N-0401 Oslo, Norway
关键词
electrothermal vaporization inductively coupled plasma mass spectrometry; lanthanides; ultrasonic slurry sampling; trifluoromethane; microwave digestion; biological samples;
D O I
10.1039/a801612g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method was developed for the determination of endogenous concentrations of the lanthanides in body fluids and tissues with electrothermal vaporization inductively coupled plasma mass spectrometry. The method was applied to the analysis of rat liver and human blood plasma as well as to some reference materials, viz., Oyster Tissue, Bovine Liver, Seronorm Trace Element Serum and Human Hair. Microwave digestion in high-pressure containers was used to digest the various biological materials. Analyses of undigested biological tissues by slurry sampling and body fluids by direct injection were also performed. For undigested samples, hydrogen peroxide injected directly onto the sample inside the graphite furnace was found to be effective in preventing the build-up of ash in the furnace. Trifluoromethane (Freon-23) was used as an effective chemical modifier that lowered the vaporization temperature and the memory effect significantly for most of the lanthanides. The absolute detection limits were, with few exceptions, in the range 1-20 fg for all the lanthanides in the various biological materials, The relative detection limits varied with the level of dilution and the method used. For Seronorm Trace Element Serum, the relative detection limits were in the range 0.05-1.2 ng l(-1). The analytical results for slurry and digested samples of the same material were similar with respect to accuracy and precision. The intermediate precision, i.e., the overall relative standard deviation obtained for the CRMs analysed on three different (not consecutive) days, was less than 25% for the lanthanides with concentrations higher than 1 ng g(-1) dry material. In the few cases where a certified or indicative value existed for the concentrations of the lanthanides in the reference materials, good agreement between these concentrations and the concentrations obtained in the present work was found. Blood plasma samples from 30 healthy volunteers and liver samples from ten unexposed rats were analysed, All the lanthanides were above the detection limit in the various samples analysed.
引用
收藏
页码:1039 / 1049
页数:11
相关论文
共 19 条
[1]  
ALLAIN P, 1990, CLIN CHEM, V36, P2012
[2]   Inductively coupled plasma mass spectrometry for direct multi-element analysis of diluted human blood and serum [J].
Barany, E ;
Bergdahl, IA ;
Schutz, A ;
Skerfving, S ;
Oskarsson, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (09) :1005-1009
[3]   RARE-EARTH ELEMENT ABUNDANCE IN TISSUES AND PLASMA OF HEALTHY-SUBJECTS AND PATIENTS BY NEUTRON-ACTIVATION ANALYSIS [J].
COLLECCHI, P ;
ESPOSITO, M ;
MELONI, S ;
ODDONE, M .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1987, 112 (02) :473-479
[4]   THE ROLE OF CHEMICAL MODIFIERS IN ANALYTE TRANSPORT LOSS INTERFERENCES WITH ELECTROTHERMAL VAPORIZATION ICP-MASS SPECTROMETRY [J].
EDIGER, RD ;
BERES, SA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (07) :907-922
[5]  
ESPOSITO M, 1986, CLIN CHEM, V32, P1598
[6]   PLASMA AND TISSUE LEVELS OF SOME LANTHANIDE ELEMENTS IN MALIGNANT AND NON-MALIGNANT HUMAN TISSUES. [J].
Esposito, Mauro ;
Collecchi, Paola ;
Brera, Silvio ;
Mora, Enzo ;
Mazzucotelli, Ambrogio ;
Cutolo, Maurizio ;
Oddone, Massimo .
Science of the Total Environment, 1986, 50 :55-63
[7]   Determination of rare earth elements in blood serum reference sample by chelating resin preconcentration and inductively coupled plasma mass spectrometry [J].
Fujimori, E ;
Tomosue, Y ;
Haraguchi, H .
TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 1996, 178 (01) :63-74
[8]   MULTIELEMENT ANALYSIS OF BIOLOGICAL TISSUES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - HEALTHY SPRAGUE-DAWLEY RATS [J].
GELINAS, Y ;
YOULA, M ;
BELIVEAU, R ;
SCHMIT, JP ;
FERRARIS, J .
ANALYTICA CHIMICA ACTA, 1992, 269 (01) :115-122
[9]   Mechanism of vaporization of yttrium and rare earth elements in electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Goltz, DM ;
Gregoire, DC ;
Chakrabarti, CL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (11) :1365-1382
[10]   A NEW METHOD FOR DETERMINATION OF RARE-EARTH ELEMENTS VAPORIZED IN GRAPHITE-FURNACE WITH A POLYTETRAFLUOROETHYLENE SLURRY FLUORINATING REAGENT BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
HUANG, M ;
JIANG, ZC ;
ZENG, Y .
ANALYTICAL SCIENCES, 1991, 7 (05) :773-778