New quantitative parameters on a recently introduced automated blood cell counter -: the XE 2100™
被引:84
作者:
Briggs, C
[1
]
Harrison, P
[1
]
Grant, D
[1
]
Staves, J
[1
]
Machin, SJ
[1
]
机构:
[1] UCL Hosp, Dept Haematol, London WC1E 6AU, England
来源:
CLINICAL AND LABORATORY HAEMATOLOGY
|
2000年
/
22卷
/
06期
关键词:
automated blood cell counter;
nucleated red cell count;
immature granulocyte count;
platelet counts;
flow cytometry;
D O I:
10.1046/j.1365-2257.2000.00330.x
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
The XE 2100 (Sysmex Corporation) is a cell counter that furthers the technology of fluorescent flow cytometry developed from the earlier range of Sysmex analysers. The new diagnostic features are a nucleated red cell count (NRBC), the ability to measure platelets by impedance as well as an 'optical' platelet count using a fluorescence dye and an immature granulocyte (IG) count, The NRBC count was highly correlated (r = 0.97) with the manual reference count. For counts below 100 x 10(9)/l the 'optical' method and the immunocount gave good a correlation (r = 0.97) optical and impedance counts were also well correlated (r = 0.89), The use of the 'optical' platelet count significantly improves the reliability of low platelet counts, The IG count correlated with visual counts (r = 0.81) and allows the detection of immature cells at an earlier stage in the laboratory process. The introduction of fluorescent flow cytometric analysis allows extended quantification of additional cell populations and so potentially improves screening and monitoring of various pathological conditions.