Computerised volumetric analysis of lesions in multiple sclerosis using new semi-automatic segmentation software

被引:13
作者
Dastidar, P
Heinonen, T
Vahvelainen, T
Elovaara, I
Eskola, H
机构
[1] Tampere Univ Hosp, Dept Diagnost Radiol, Tampere, Finland
[2] Tampere Univ Technol, Ragnar Granit Inst, FIN-33101 Tampere, Finland
[3] Tampere Univ Hosp, Dept Neurol & Rehabil, Tampere, Finland
[4] Tampere Univ, Sch Med, SF-33520 Tampere, Finland
关键词
multiple sclerosis; magnetic resonance; image processing; volume measurement; segmentation; 3D visualisation;
D O I
10.1007/BF02513274
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The paper describes the application of new semi-automatic segmentation software to the task of detection of anatomical structures and lesion and their three-dimensional (3D) visualisation in 23 patients with secondary progressive multiple sclerosis (MS). The purpose is to study the correlation between magnetic resonance imaging (MRI) parameters (volumes of plaques and cerebrospinal fluid spaces) and clinical deficits (neurological deficits in the form of EDSS and RFSS scores, and neuropsychological deficits). The software operates in PC/Windows and PC/NeXTstep environments and utilises graphical user interfaces. Quantitative accuracy is measured by performing segmentation of fluid-filled syringes (relative error of 1.5%), and reproducibility is measured by intra- and inter-observer studies (3% and 7% variability, respectively). The mean volumes of MS plaques show significant correlations with the total RFSS scores (p=0.04). Relative intracranial cerebrospinal fluid (CSF) space volumes show statistically significant correlation with EDSS scores (p=0.01). The mean volume of MS plaques shows a significant correlation with the overall neuropsychological deficits (p=0.03). 3D visualisation helps to understand the relationship of lesions to the surrounding brain structures. The use of semiautomatic segmentation techniques is recommended in the clinical diagnosis of MS patients.
引用
收藏
页码:104 / 107
页数:4
相关论文
共 18 条
[1]   CLINICALLY ISOLATED LESIONS OF THE TYPE SEEN IN MULTIPLE-SCLEROSIS - A COGNITIVE, PSYCHIATRIC, AND MRI FOLLOW-UP-STUDY [J].
FEINSTEIN, A ;
KARTSOUNIS, LD ;
MILLER, DH ;
YOUL, BD ;
RON, MA .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1992, 55 (10) :869-876
[2]   Interscanner variation in brain MRI lesion load measurements in MS: Implications for clinical trials [J].
Filippi, M ;
vanWaesberghe, JH ;
Horsfield, MA ;
Bressi, S ;
Gasperini, C ;
Yousry, TA ;
GawneCain, ML ;
Morrissey, SP ;
Rocca, MA ;
Barkhof, F ;
Nijeholt, GJLA ;
Bastianello, S ;
Miller, DH .
NEUROLOGY, 1997, 49 (02) :371-377
[3]  
Filippi M, 1996, ACTA NEUROL SCAND, V93, P30
[4]   Tumour volume determination from MR images by morphological segmentation [J].
Gibbs, P ;
Buckley, DL ;
Blackband, SJ ;
Horsman, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2437-2446
[5]  
GONSALEZ CF, 1994, J NEUROIMAGING, V4, P188
[6]   Quantification of MRI lesion load in multiple sclerosis: A comparison of three computer-assisted techniques [J].
Grimaud, J ;
Lai, M ;
Thorpe, J ;
Adeleine, P ;
Wang, L ;
Barker, GJ ;
Plummer, DL ;
Tofts, PS ;
McDonald, WI ;
Miller, DH .
MAGNETIC RESONANCE IMAGING, 1996, 14 (05) :495-505
[7]   Applicability of semi-automatic segmentation for volumetric analysis of brain lesions [J].
Heinonen, T ;
Dastidar, P ;
Eskola, H ;
Frey, H ;
Ryymin, P ;
Laasonen, E .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1998, 22 (04) :173-178
[8]   3D visualization library for multimodal medical images [J].
Heinonen, T ;
Visala, K ;
Blomqvist, M ;
Eskola, H ;
Frey, H .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1998, 22 (04) :267-273
[9]   Segmentation of T1 MR scans for reconstruction of resistive head models [J].
Heinonen, T ;
Eskola, H ;
Dastidar, P ;
Laarne, P ;
Malmivuo, J .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1997, 54 (03) :173-181
[10]   Semi-automatic tool for segmentation and volumetric analysis of medical images [J].
Heinonen, T ;
Dastidar, P ;
Kauppinen, P ;
Malmivuo, J ;
Eskola, H .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (03) :291-296