Accuracy of the Enlite 6-Day Glucose Sensor with Guardian and Veo Calibration Algorithms

被引:66
作者
Keenan, Desmond Barry [1 ]
Mastrototaro, John Joseph [1 ]
Zisser, Howard [2 ]
Cooper, Kenneth A. [1 ]
Raghavendhar, Gautham [1 ]
Lee, Scott W. [1 ]
Yusi, Jonathan [1 ]
Bailey, Timothy S. [3 ]
Brazg, Ronald Leonard [4 ]
Shah, Rajiv V. [1 ]
机构
[1] Medtron Diabet, R&D, Northridge, CA 91325 USA
[2] Samsung Diabet Res Ctr, Santa Barbara, CA USA
[3] Univ Calif San Diego, Sch Med, San Diego, CA 92103 USA
[4] Rainer Clin Res Ctr Inc, Renton, WA USA
关键词
IMPROVED GLYCEMIC CONTROL; INSULIN-PUMP THERAPY; BLOOD-GLUCOSE; SYSTEM;
D O I
10.1089/dia.2011.0199
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Objective: This study investigates the accuracy of a newly developed, next-generation subcutaneous glucose sensor, evaluated for 6-day use. Research Design and Methods: Seventy-nine subjects (53 men, 26 women) with type 1 diabetes and 18 subjects (14 men, four women) with type 2 diabetes completed a three-center, prospective, sensor accuracy study. The mean age for the group was 42.2 +/- 15.0 years (mean +/- SD), ranging from 18 to 71 years, with a mean glycosylated hemoglobin level of 7.6 +/- 1.5%, ranging from 5.5% to 14%. Subjects wore Enlite(TM) sensors (Medtronic Diabetes, Northridge, CA) in the abdominal and buttocks region for two separate 7-day periods and calibrated with a home-use blood glucose meter. Subjects participated in an in-clinic testing day where frequent sampled plasma glucose samples were acquired every 15 min for 10 h. Sensor data was retrospectively processed with Guardian(R) REAL-Time (Medtronic) and Paradigm(R) Veo(TM) (Medtronic) calibration routines, and accuracy metrics were calculated for each algorithm and sensor location. Physiological time lag for each measurement site was calculated. Results: Based on 6,404 plasma-sensor glucose paired points, the Enlite sensor with Veo calibration algorithm produced a mean absolute relative difference of 13.86% with 97.3% of points within the A+B zones of the Clarke error grid. Threshold-only alarms detected 90.1% of hypoglycemia and 90% of hyperglycemia. Mean time lag measured at the abdominal region was 7.94 +/- 6.48 min compared with 11.70 +/- 6.71 min (P < 0.0001) at the buttocks area. Conclusions: The Enlite sensor accurately measures glucose when compared with gold standard laboratory measurements over its 6-day use. Sensors placed in the buttocks region exhibited greater time lags than sensors placed in the abdomen.
引用
收藏
页码:225 / 231
页数:7
相关论文
共 24 条
[1]
[Anonymous], POCT05A CLSI
[2]
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes [J].
Bergenstal, Richard M. ;
Tamborlane, William V. ;
Ahmann, Andrew ;
Buse, John B. ;
Dailey, George ;
Davis, Stephen N. ;
Joyce, Carol ;
Peoples, Tim ;
Perkins, Bruce A. ;
Welsh, John B. ;
Willi, Steven M. ;
Wood, Michael A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (04) :311-320
[3]
Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes [J].
Castle, Jessica R. ;
Engle, Julia M. ;
Youssef, Joseph E. L. ;
Massoud, Ryan G. ;
Yuen, Kevin C. J. ;
Kagan, Ryland ;
Ward, W. Kenneth .
DIABETES CARE, 2010, 33 (06) :1282-1287
[4]
EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[5]
Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring [J].
Deis, Dorothee ;
Bolinder, Jan ;
Riveline, Jean-Pierre ;
Battelino, Tadej ;
Bosi, Emanuele ;
Tubiana-Rufi, Nadia ;
Kerr, David ;
Phillip, Moshe .
DIABETES CARE, 2006, 29 (12) :2730-2732
[6]
Continuous home monitoring of glucose - Improved glycemic control with real-life use of continuous glucose sensors in adult subjects with type 1 diabetes [J].
Garg, Satish K. ;
Kelly, William C. ;
Voelmle, Mary K. ;
Ritchie, Peter J. ;
Gottlieb, Peter A. ;
McFann, Kim K. ;
Ellis, Samuel L. .
DIABETES CARE, 2007, 30 (12) :3023-3025
[7]
Sensor-augmented pump therapy lowers HbA1c in suboptimally controlled Type 1 diabetes; a randomized controlled trial [J].
Hermanides, J. ;
Norgaard, K. ;
Bruttomesso, D. ;
Mathieu, C. ;
Frid, A. ;
Dayan, C. M. ;
Diem, P. ;
Fermon, C. ;
Wentholt, I. M. E. ;
Hoekstra, J. B. L. ;
DeVries, J. H. .
DIABETIC MEDICINE, 2011, 28 (10) :1158-1167
[8]
Sensor-augmented insulin pump therapy: Results of the first randomized treat-to-target study [J].
Hirsch, Irl B. ;
Abelseth, Jill ;
Bode, Bruce W. ;
Fischer, Jerome S. ;
Kaufman, Francine R. ;
Mastrototaro, John ;
Parkin, Christopher G. ;
Wolpert, Howard A. ;
Buckingham, Bruce A. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2008, 10 (05) :377-383
[9]
Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial [J].
Hovorka, Roman ;
Allen, Janet M. ;
Elleri, Daniela ;
Chassin, Ludovic J. ;
Harris, Julie ;
Xing, Dongyuan ;
Kollman, Craig ;
Hovorka, Tomas ;
Larsen, Anne Mette F. ;
Nodale, Marianna ;
De Palma, Alessandro ;
Wilinska, Malgorzata E. ;
Acerini, Carlo L. ;
Dunger, David B. .
LANCET, 2010, 375 (9716) :743-751
[10]
International Organization for Standardization, 2003, ISO 15197 IN VITR DI