Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery

被引:41
作者
Godje, O [1 ]
Hoke, K [1 ]
Lamm, P [1 ]
Schmitz, C [1 ]
Thiel, C [1 ]
Weinert, M [1 ]
Reichart, B [1 ]
机构
[1] Univ Munich, Univ Hosp Grosshadern, Dept Cardiac Surg, D-8000 Munich, Germany
关键词
thermodilution; cardiac output; cardiac surgery; pulse contour analysis; intensive care medicine;
D O I
10.1055/s-2007-1010233
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A pulse-contour-based method for continuous measurement of cardiac output (CO) and systemic vascular resistance (SVR) was tested and arterial thermodilution, used for calibration, was compared to pulmonary artery thermodilution. In 30 patients CO and SVR were measured by pulse contour analysis (COpc, SVRpc) 270 times in 24 h and compared to arterial (COart, SVRart) and pulmonary arterial (COpa, SVRpa) thermodilution measurements. The mean difference between COpa and COart was 0.26 L/min (3.6%) with a standard deviation (SD) of 0.7 L/min, the correlation coefficient was 0.96, and the coefficient of variation was 5.0% and 5.9% respectively. COpc did differ from COpa by 0.11 L/min (1.5%, SD = 0.6 L/min) and from COart by 0.15 L/min (2.1 %, SD = 0.7 L/min). Correlation of COpc with COpa was 0.91, correlation of COpc with COart was 0.90. SVRpc did correlate with SVRpa, a coefficient of 0.94, and with SVRart, a coefficient of 0.92. Mean COpc and SVRpc did not differ significantly from COpa or COart and SVRpa or SVRart during the 24 h study period. It is concluded that COart correlates well with COpa and can be used to calibrate COpc. COpc and SVRpc agree with thermodilution-based CO and SVR without recalibration for 24 hours.
引用
收藏
页码:242 / 249
页数:8
相关论文
共 20 条
[1]   CONSERVATION OF THERMAL INDICATOR IN LUNG CIRCULATION [J].
ARFORS, K ;
MALMBERG, P ;
PAVEK, K .
CARDIOVASCULAR RESEARCH, 1971, 5 (04) :530-+
[2]   CARDIAC-OUTPUT MEASUREMENT USING FEMORAL-ARTERY THERMODILUTION IN PATIENTS [J].
BOCK, JC ;
BARKER, BC ;
MACKERSIE, RC ;
TRANBAUGH, RF ;
LEWIS, FR .
JOURNAL OF CRITICAL CARE, 1989, 4 (02) :106-111
[3]   RELATIONSHIP BETWEEN CO AND TRANSIT TIMES FOR DYE AND THERMAL INDICATORS IN CENTRAL CIRCULATION [J].
CARLILE, PV ;
BECKETT, RC ;
GRAY, BA .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 60 (04) :1363-1372
[4]   The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[5]   Is it time to pull the pulmonary artery catheter? [J].
Dalen, JE ;
Bone, RC .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :916-918
[6]  
GODJE O, 1997, CRIT CARE MED S1, V25, P50
[7]  
GORE JM, 1987, CHEST, V98, P1331
[8]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[9]   THE SLOWING OF SINUS RHYTHM DURING THERMODILUTION CARDIAC-OUTPUT DETERMINATION AND THE EFFECT OF ALTERING INJECTATE TEMPERATURE [J].
HARRIS, AP ;
MILLER, CF ;
BEATTIE, C ;
ROSENFELD, GI ;
ROGERS, MC .
ANESTHESIOLOGY, 1985, 63 (05) :540-541
[10]   CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT USING PULSE CONTOUR ANALYSIS [J].
IRLBECK, M ;
FORST, H ;
BRIEGEL, J ;
HALLER, M ;
PETER, K .
ANAESTHESIST, 1995, 44 (07) :493-500