Comparison of the eye's wave-front aberration measured psychophysically and with the Shack-Hartmann wave-front sensor

被引:91
作者
Salmon, TO [1 ]
Thibos, LN
Bradley, A
机构
[1] NE State Univ, Coll Optometry, Tahlequah, OK 74464 USA
[2] Indiana Univ, Sch Optometry, Bloomington, IN 47405 USA
来源
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION | 1998年 / 15卷 / 09期
关键词
D O I
10.1364/JOSAA.15.002457
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The Shack-Hartmann wave-front sensor offers many theoretical advantages over other methods for measuring aberrations of the eye; therefore it is essential that its accuracy be thoroughly tested. We assessed the accuracy of a Shack-Hartmann sensor by directly comparing its measured wave-front aberration function with that obtained by the Smirnov psychophysical method for the same eyes. Wave-front profiles measured by the two methods agreed closely in terms of shape and magnitude with rms differences of -lambda/2 and -lambda/6 (5.6-mm pupil) for two eyes. Primary spherical aberration was dominant in these profiles, and, in one subject, secondary coma was opposite in sign to primary coma, thereby canceling its effect. Discovery of an unusual, subtle wave-front anomaly in one individual further demonstrated the accuracy and sensitivity of the Shack-Hartmann wave-front sensor for measuring the optical quality of the human eye. (C) 1998 Optical Society of America.
引用
收藏
页码:2457 / 2465
页数:9
相关论文
共 25 条
[1]   MEASUREMENT OF MONOCHROMATIC OCULAR ABERRATIONS OF HUMAN EYES AS A FUNCTION OF ACCOMMODATION BY THE HOWLAND ABERROSCOPE TECHNIQUE [J].
ATCHISON, DA ;
COLLINS, MJ ;
WILDSOET, CF ;
CHRISTENSEN, J ;
WATERWORTH, MD .
VISION RESEARCH, 1995, 35 (03) :313-323
[2]  
Born M., 1986, PRINCIPLES OPTICS
[3]   WAVE-FRONT ABERRATION OF THE EYE - A REVIEW [J].
CHARMAN, WN .
OPTOMETRY AND VISION SCIENCE, 1991, 68 (08) :574-583
[4]  
Goodman J.W., 1996, Opt. Eng, V35, P1513, DOI DOI 10.1016/J.APSUSC.2017.08.033
[5]   A REFLECTOMETRIC TECHNIQUE FOR ASSESSING PHOTORECEPTOR ALIGNMENT [J].
GORRAND, JM ;
DELORI, F .
VISION RESEARCH, 1995, 35 (07) :999-1010
[6]   SUBJECTIVE METHOD FOR MEASUREMENT OF MONOCHROMATIC ABERRATIONS OF EYE [J].
HOWLAND, HC ;
HOWLAND, B .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1977, 67 (11) :1508-1518
[7]   Supernormal vision and high-resolution retinal imaging through adaptive optics [J].
Liang, JZ ;
Williams, DR ;
Miller, DT .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (11) :2884-2892
[8]   OBJECTIVE MEASUREMENT OF WAVE ABERRATIONS OF THE HUMAN EYE WITH THE USE OF A HARTMANN-SHACK WAVE-FRONT SENSOR [J].
LIANG, JZ ;
GRIMM, B ;
GOELZ, S ;
BILLE, JF .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1994, 11 (07) :1949-1957
[9]   Aberrations and retinal image quality of the normal human eye [J].
Liang, JZ ;
Williams, DR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (11) :2873-2883
[10]  
Mandell Robert B., 1994, CLAO Journal, V20, P123