Use of the Less Invasive Stabilization System (LISS) in patients with distal femoral (AO33) fractures:: a prospective multicenter study

被引:169
作者
Schütz, M
Müller, M
Regazzoni, P
Höntzsch, D
Krettek, C
Van der Werken, C
Haas, N
机构
[1] AO Clin Invest & Documentat, CH-8600 Dubendorf, Switzerland
[2] Virchow Klinikum Humboldt Univ Berlin, Klin Unfall & Wiederherstellungschirurg, D-13353 Berlin, Germany
[3] Kantonsspital, CH-4031 Basel, Switzerland
[4] BG Unfallklin, D-72076 Tubingen, Germany
[5] Hannover Med Sch, Trauma Dept Hannover, D-30625 Hannover, Germany
[6] Acad Ziekenhuis Utrecht, NL-3508 GA Utrecht, Netherlands
关键词
femoral fractures; internal fixation device; internal fracture fixation; leg injury;
D O I
10.1007/s00402-004-0779-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction The Less Invasive Stabilization System (LISS) is an internal fixator that can be inserted percutaneously by means of a minimally invasive surgical approach. This paper presents the use of this system exclusively in patients with AO33 distal femoral fractures, a group in which fracture management is often complicated by multiple trauma and soft-tissue damage. Materials and methods A series of 62 patients (mean age 52 years) with 66 AO33A or C fractures received the LISS implant and were followed prospectively for 12 months. Assessments included the radiographic evaluation of implant placement and healing, measurement of valgus/varus and sagittal joint malalignment, and the assessment of weight-bearing capacity and range of motion. Results The mean duration of surgery was 85 min (range 40-135 min) for 33A fractures and 149 min (range 50-300 min) for 33C fractures. Postoperative radiographic assessments showed that the LISS implant was positioned correctly in 59 cases (89%). Assessment of valgus/varus alignment showed correct axial alignment in 49 cases (74%), a deviation of 5-10 degrees in 13 cases, and a 10-20 degrees deviation in 1 case. Correct sagittal alignment was observed in 56 cases (85%), 5-10 degrees malalignment in 5 cases, and 10-20 degrees in 2 cases. During the course of the 1-year follow-up, 8 patients (9 fractures) died; 2 other patients were not available for follow-up. Complete fracture healing was achieved in 85% of the followed-up patients. Forty-eight patients were capable of full weight-bearing. After fracture healing, maximum flexion of the injured limb was on average 80% of the range of the uninjured limb. An extension deficit of more than 5 degrees was measured in 3 cases. Of the 62 patients, 14 underwent further operations during the course of follow-up. Of these, 6 required bone grafting and 3 refixation of the implant due to implant loosening. Deep infections requiring several debridements occurred in 2 patients. No complications relating solely to the implant were observed. Conclusion The LISS showed good overall results in the treatment of these difficult fractures. It is a good alternative to conventional extramedullary and intramedullary stabilizing techniques, especially in more complex fracture situations.
引用
收藏
页码:102 / 108
页数:7
相关论文
共 24 条
[1]
BUHREN V, 1987, UNFALLCHIRURGIE, V3, P152
[2]
DANZIGER MB, 1995, AM J ORTHOP, V9, P684
[3]
*EUR COMM STAND, 1993, 540 EN EUR COMM STAN, P1
[4]
Femur Kinzl L., 2000, AO PRINCIPLES FRACTU
[5]
FRIGG R, 2001, INBJURY, pSC24
[6]
Clinical trial of the distal femur nail [J].
Grass, R ;
Biewener, A ;
Endres, T ;
Rammelt, S ;
Barthel, S ;
Zwipp, H .
UNFALLCHIRURG, 2002, 105 (09) :783-790
[7]
PROBLEMS IN THE MANAGEMENT OF TYPE-III (SEVERE) OPEN FRACTURES - A NEW CLASSIFICATION OF TYPE-III OPEN FRACTURES [J].
GUSTILO, RB ;
MENDOZA, RM ;
WILLIAMS, DN .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1984, 24 (08) :742-746
[8]
GUSTILO RB, 1982, INSTR COURS LECT, P64
[9]
Indirect reduction and bridge plating of supracondylar fractures of the femur [J].
Huang, HT ;
Huang, PJ ;
Su, JY ;
Lin, SY .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 (02) :135-140
[10]
INITIAL EXPERIENCE WITH THE TREATMENT OF SUPRACONDYLAR FEMORAL FRACTURES USING THE SUPRACONDYLAR INTRAMEDULLARY NAIL - A PRELIMINARY-REPORT [J].
IANNACONE, WM ;
BENNETT, FS ;
DELONG, WG ;
BORN, CT ;
DALSEY, RM .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1994, 8 (04) :322-327