Characterization of a noncolinear acousto-optic tunable filter and its resolution enhancement as a near-infrared spectrometer

被引:14
作者
Cheng, AYS [1 ]
Zhu, JH
Pau, MCY
机构
[1] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon, Hong Kong, Peoples R China
[2] Sichuan Univ, Informat Opt Inst, Chengdu 610064, Peoples R China
关键词
acousto-optic tunable filter; near-infrared spectroscopy; Fourier deconvolution; resolution enhancement;
D O I
10.1366/0003702011951768
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An automatic near-infrared (NIR) spectrometer based on a non-colinear acousto-optic tunable filter (AOTF) is constructed for highspeed spectrometry. Its initial spectral characteristics are studied systematically for application. The AOTF is a solid-state, rugged, compact, and high-speed device and can be readily used for spectral measurement applications, but some broadening and overlapping effects will occur when it is used for the measurement of a narrowband spectrum. In order to correct the spectral distortion and obtain a fine spectrum, a new resolution enhancement procedure, which combines conventional Fourier self-deconvolution (FSD) with our optimized evaluation function and objective parameters determination, is used to resolve the band-broadened and band-overlapped spectral lines. The self-deconvolved spectrum is in good agreement with the fine spectrum measured by a high-resolution grating spectrometer. This is, to our knowledge, the first resolution enhancement of an AOTF spectrum by analytical and numerical operation.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 15 条
[1]   High-resolution spectroscopy using an acousto-optic tunable filter and a fiber-optic Fabry-Perot interferometer [J].
Baldwin, DP ;
Zamzow, DS ;
DSilva, AP .
APPLIED SPECTROSCOPY, 1996, 50 (04) :498-503
[2]   Characterization of an acousto-optic tunable filter and use in visible spectrophotometry [J].
Bucher, EG ;
Carnahan, JW .
APPLIED SPECTROSCOPY, 1999, 53 (05) :603-611
[3]   A GENERALIZED-APPROACH TO DERIVATIVE SPECTROSCOPY [J].
CAMERON, DG ;
MOFFATT, DJ .
APPLIED SPECTROSCOPY, 1987, 41 (04) :539-544
[4]  
Chang I. C., 1976, Proceedings of the Society of Photo-Optical Instrumentation Engineers. Vol.90. Device Development/Instrumentation/Applications, P12
[5]   NONCOLLINEAR ACOUSTOOPTIC FILTER WITH LARGE ANGULAR APERTURE [J].
CHANG, IC .
APPLIED PHYSICS LETTERS, 1974, 25 (07) :370-372
[6]   FLOW-RATE MEASUREMENTS IN ISOLATED PERFUSED KIDNEY-TUBULES BY FLUORESCENCE PHOTOBLEACHING RECOVERY [J].
FLAMION, B ;
BUNGAY, PM ;
GIBSON, CC ;
SPRING, KR .
BIOPHYSICAL JOURNAL, 1991, 60 (05) :1229-1242
[7]   ACOUSTO-OPTIC TUNABLE FILTER [J].
HARRIS, SE ;
WALLACE, RW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1969, 59 (06) :744-+
[8]  
HAYKIN S, 1979, TOPICS APPL PHYSICS, V34
[9]   FOURIER SELF-DECONVOLUTION - A METHOD FOR RESOLVING INTRINSICALLY OVERLAPPED BANDS [J].
KAUPPINEN, JK ;
MOFFATT, DJ ;
MANTSCH, HH ;
CAMERON, DG .
APPLIED SPECTROSCOPY, 1981, 35 (03) :271-276
[10]   NOISE IN FOURIER SELF-DECONVOLUTION [J].
KAUPPINEN, JK ;
MOFFATT, DJ ;
CAMERON, DG ;
MANTSCH, HH .
APPLIED OPTICS, 1981, 20 (10) :1866-1879