Morphological characterization and in vitro biocompatibility of a porous nickel-titanium alloy

被引:92
作者
Prymak, O
Bogdanski, D
Köller, M
Esenwein, SA
Muhr, G
Beckmann, F
Donath, T
Assad, M
Epple, M
机构
[1] Univ Duisburg, Inst Inorgan Chem, D-45117 Essen, Germany
[2] BG Kliniken Bergmannsheil, Dept Surg, D-44789 Bochum, Germany
[3] GKSS Forschungszentrum Geesthacht GmbH, Mat Res Inst, D-25102 Geesthacht, Germany
[4] McGill Univ, Ctr Bone & Periodon Res, Montreal, PQ H3A 1A4, Canada
关键词
shape memory alloys; biocompatibility; cytokines; leukocytes; porosity; nickel-titanium alloys;
D O I
10.1016/j.biomaterials.2005.02.029
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Disks consisting of macroporous nickel-titanium alloy (NiTi, Nitinol (R), Actipore (R)) are used as implants in clinical surgery, e.g. for fixation of spinal dysfunctions. The morphological properties were studied by scanning electron microscopy (SEM) and by synchrotron radiation-based microtomography (SR mu CT). The composition was studied by X-ray diffractometry (XRD), differential scanning calorimetry (DSC), and energy-dispersive X-ray spectroscopy (EDX). The mechanical properties were studied with temperature-dependent dynamical mechanical analysis (DMA). Studies on the biocompatibility were performed by co-incubation of porous NiTi samples with isolated peripheral blood leukocyte fractions (polymorphonuclear neutrophil granulocytes, PMN; peripheral blood mononuclear leukocytes, PBMC) in comparison with control cultures without NiTi samples. The cell adherence to the NiTi surface was analyzed by fluorescence microscopy and scanning electron microscopy. The activation of adherent leukocytes was analyzed by measurement of the released cytokines using enzyme-linked immunosorbent assay (ELISA). The cytokine response of PMN (analyzed by the release of IL-1ra and IL-8) was not significantly different between cell cultures with or without NiTi. There was a significant increase in the release of IL-1ra (p < 0.001), IL-6 (p < 0.05), and IL-8 (p < 0.05) from PBMC in the presence of NiTi samples. In contrast, the release of TNF-alpha by PBMC was not significantly elevated in the presence of NiTi. IL-2 was released from PBMC only in the range of the lower detection limit in all cell cultures. The material, clearly macroporous with an interconnecting porosity, consists of NiTi (martensite; monoclinic, and austenite; cubic) with small impurities of NiTi2 and possibly NiCx,. The material is not superelastic upon manual compression and shows a good biocompatibility. (c) 2005 Elsevier Ltd. All rights reserved.
引用
收藏
页码:5801 / 5807
页数:7
相关论文
共 40 条
[1]   COMPARISON OF BENDING AND TENSION TESTS FOR ORTHODONTIC WIRES [J].
ASGHARNIA, MK ;
BRANTLEY, WA .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1986, 89 (03) :228-236
[2]   Porous nitinol vs. titanium intervertebral fusion implants: Computer tomography, radiological and histological study of osseointegration capacity [J].
Assad, M ;
Likibi, F ;
Jarzem, P ;
Leroux, MA ;
Coillard, C ;
Rivard, CH .
MATERIALWISSENSCHAFT UND WERKSTOFFTECHNIK, 2004, 35 (04) :219-223
[3]   Porous titanium-nickel for intervertebral fusion in a sheep model: Part 2. Surface analysis and nickel release assessment [J].
Assad, M ;
Chernyshov, AV ;
Jarzem, P ;
Leroux, MA ;
Coillard, C ;
Charette, S ;
Rivard, CH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :121-129
[4]   Porous titanium-nickel for intervertebral fusion in a sheep model: Part 1. Histomorphometric and radiological analysis [J].
Assad, M ;
Jarzem, P ;
Leroux, MA ;
Coillard, C ;
Chernyshov, AV ;
Charette, S ;
Rivard, CH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :107-120
[5]  
Assad M, 1998, J BIOMED MATER RES, V41, P154, DOI 10.1002/(SICI)1097-4636(199807)41:1<154::AID-JBM18>3.0.CO
[6]  
2-N
[7]  
BAROUK L S, 1992, Journal of Foot Surgery, V31, P388
[8]  
BAROUK LS, 1994, FOOT DIS, V1, P79
[9]   New developments in attenuation and phase-contrast microtomography using synchrotron radiation with low and high photon energies [J].
Beckmann, F ;
Bonse, U ;
Biermann, T .
DEVELOPMENTS IN X-RAY TOMOGRAPHY II, 1999, 3772 :179-187
[10]   Biocompatibility of calcium phosphate-coated and of geometrically structured nickel-titanium (NiTi) by in vitro testing methods [J].
Bogdanski, D ;
Epple, M ;
Esenwein, SA ;
Muhr, G ;
Petzoldt, V ;
Prymak, O ;
Weinert, K ;
Köller, M .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 2004, 378 (1-2) :527-531