Measurement of carboxyhemoglobin in forensic blood samples using UV-visible spectrometry and improved principal component regression

被引:41
作者
Egan, WJ
Brewer, WE
Morgan, SL [1 ]
机构
[1] Univ S Carolina, Dept Chem & Biochem, Columbia, SC 29208 USA
[2] S Carolina Law Enforcement Div, Dept Toxicol, Columbia, SC 29210 USA
关键词
chemometrics; principal component regression; PCR; forensic chemistry; UV-vis spectrometry; carboxyhemoglobin; carbon monoxide; confidence and prediction intervals;
D O I
10.1366/0003702991946352
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The forensic determination of carboxyhemoglobin (COHb) in blood was performed by using an improved principal component regression (PCR) technique applied to UV-visible spectra. Calibration data were decomposed into principal components, and the principal components useful for prediction were selected by their correlation with calibration spectra. Cross-validation of prediction results was done by leverage-corrected residuals. Confidence and prediction intervals derived from classical regression theory were found to be reasonable in size. The results compared favorably to a comparison study conducted by using a CO Oximeter method. In analysis of forensic case study samples, the improved PCR method allowed detection of abnormal samples and successfully predicted percentages of COHb and methemoglobin (MetHb), and provided error estimates for those predictions.
引用
收藏
页码:218 / 225
页数:8
相关论文
共 61 条
[31]  
KLENDSHOJ NC, 1950, J BIOL CHEM, V183, P297
[32]   DIRECT CARBOXYHEMOGLOBIN DETERMINATION BY DERIVATIVE SPECTROSCOPY [J].
LOPEZRIVADULLA, M ;
BERMEJO, AM ;
FERNANDEZ, P ;
CRUZ, A ;
CONCHEIRO, L .
FORENSIC SCIENCE INTERNATIONAL, 1989, 40 (03) :261-266
[33]   ALTERNATIVES TO CROSS-VALIDATORY ESTIMATION OF THE NUMBER OF FACTORS IN MULTIVARIATE CALIBRATION [J].
LORBER, A ;
KOWALSKI, BR .
APPLIED SPECTROSCOPY, 1990, 44 (09) :1464-1470
[34]  
MAHONEY JJ, 1993, CLIN CHEM, V39, P1693
[35]   USE OF THE SINGULAR VALUE DECOMPOSITION IN REGRESSION-ANALYSIS [J].
MANDEL, J .
AMERICAN STATISTICIAN, 1982, 36 (01) :15-24
[36]  
MARTENS H, 1987, NEAR INFRARED TECHNO, pCH4
[37]  
Martens H., 1989, Multivariate Calibration
[38]  
MOSTELLER F, 1986, COLLECTED WORKS JW T, V4, pCH15
[39]  
NAES T, 1985, TECHNOMETRICS, V27, P301