MR-guided placement of a temporary vena cava filter: Technique and feasibility

被引:23
作者
Frahm, C [1 ]
Gehl, HB [1 ]
Lorch, H [1 ]
Zwaan, M [1 ]
Drobnitzky, M [1 ]
Laub, GA [1 ]
Weiss, HD [1 ]
机构
[1] Univ Lubeck, Dept Radiol, D-23538 Lubeck, Germany
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 01期
关键词
interventional MRI; vascular interventions; vena cava filter; passive visualization; low-field MRI;
D O I
10.1002/jmri.1880080121
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this paper was to investigate the feasibility of MR-guided insertion of a temporary vena cava Biter on an open low-field imager, In vivo procedures were performed on four anesthetized pigs using a common nonferromagnetic temporary vena cava filter and a special prototype guidewire developed for vascular interventions guided by low-field MRI, Breath-hold spoiled gradient-echo sequences (fast low-angle shot [FLASH]) with now compensation were used for position monitoring of the passively visualized intravascular devices. Using the described technique and equipment, all steps of the procedure were feasible in the MR unit. Practicability of the procedure seems to be sufficient for clinical purposes but was inferior compared to the conventional technique of filter placement.
引用
收藏
页码:105 / 109
页数:5
相关论文
共 19 条
[1]  
Frahm C, 1996, CARDIOVASC INTER RAD, V19, P335
[2]  
FRAHM C, 1997, P 5 SCI M INT SOC MA
[3]   THE NITINOL,SIMON FILTER - EVALUATION BY MR AND ULTRASOUND [J].
KIM, D ;
EDELMAN, RR ;
MARGOLIN, CJ ;
PORTER, DH ;
MCARDLE, CR ;
SCHLAM, BW ;
GIANTURCO, LE ;
SIEGEL, JB ;
SIMON, M .
ANGIOLOGY, 1992, 43 (07) :541-548
[4]   VASCULAR INTERVENTIONS GUIDED BY ULTRAFAST MR-IMAGING - EVALUATION OF DIFFERENT MATERIALS [J].
KOCHLI, VD ;
MCKINNON, GC ;
HOFMANN, E ;
VONSCHULTHESS, GK .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (03) :309-314
[5]   Temporary cava filter: Effective prophylaxis of pulmonary embolism in venous thromboses in the region of the pelvic vascular system and of the vena cava inferior? [J].
Kunisch, M ;
Rauber, K ;
Bachmann, G ;
Rau, WS .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 163 (06) :523-526
[6]   Biopsy needle susceptibility artifacts [J].
Ladd, ME ;
Erhart, P ;
Debatin, JF ;
Romanowski, BJ ;
Boesiger, P ;
McKinnon, GC .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (04) :646-651
[7]  
LADD ME, 1997, P ISMRM 5 ANN M VANC, P1937
[8]   INTRAVASCULAR MR TRACKING CATHETER - PRELIMINARY EXPERIMENTAL EVALUATION [J].
LEUNG, DA ;
DEBATIN, JF ;
WILDERMUTH, S ;
MCKINNON, GC ;
HOLTZ, D ;
DUMOULIN, CL ;
DARROW, RD ;
HOFMANN, E ;
VONSCHULTHESS, GK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (05) :1265-1270
[9]   Needle localization in MR-guided biopsy and aspiration: Effects of field strength, sequence design, and magnetic field orientation [J].
Lewin, JS ;
Duerk, JL ;
Jain, VR ;
Petersilge, CA ;
Chao, CP ;
Haaga, JR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (06) :1337-1345
[10]  
LINSENMAIER U, 1996, CARDIVOASC INTERV S2, V19, pS94