Limits of preservation of samples for urine strip tests and particle counting

被引:35
作者
Kouri, Timo [1 ,2 ]
Malminiemi, Outi [3 ]
Penders, Joris [4 ,5 ]
Pelkonen, Virpi [1 ,2 ]
Vuotari, Lotta [3 ]
Delanghe, Joris [4 ]
机构
[1] Oulu Univ Hosp, Oulu, Finland
[2] Oulu Univ, Dept Clin Chem, Oulu, Finland
[3] Pirkanmaa Hosp Dist, Ctr Lab Med, Tampere, Finland
[4] Univ Ghent, Dept Clin Chem, B-9000 Ghent, Belgium
[5] Ziekenhuis Oost Limburg, Dept Biol Clin, Genk, Belgium
关键词
automation; particle counting; preservation; test strips; urinalysis;
D O I
10.1515/CCLM.2008.122
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Preservation of urine samples is important for centralised laboratory services with automated instruments. Methods: A multicentre evaluation was carried out to assess preservative tubes from BD Diagnostics-Pre-analytical Systems and from Greiner Bio-One for test strip reading (documented at the level of remission values), for particle counting by flow cytometers (UF-100) and for visual microscopy. Failures were expressed as percentages of originally positive samples beyond a two-fold change (+100% or -50%) from the original values. Results: The preservative-containing BD Plus C&S plastic, BD Plus UAP and Greiner Stabilur tubes succeeded in preservation of test strip results for 6-24 h (exceptions were glucose and nitrite tests). Greiner boric acid tube showed false negative results in leukocyte, protein and ketone strip tests immediately after adding the preservative. Urine red blood cell counts (with Sysmex UF-100) were preserved for 5 h in BID Plus C&S plastic and Greiner Stabilur tubes (Greiner tubes having clearly larger preservative-related background). Bacteria or white blood cell counting succeeded in BD Plus C&S plastic tubes for 5 or 24 h, respectively, but up to 72 h in Greiner Stabilur tubes. In visual microscopy, the Greiner Stabilur tube was slightly better than the BD Plus C&S plastic tube. Conclusions: Urine specimens can be transported at +20 degrees C on the day of collection if preserved properly. Longer delays need careful planning with current preservatives. Flow cytometry with UF-100 is sensitive to non-dissolved preservative remnants.
引用
收藏
页码:703 / 713
页数:11
相关论文
共 13 条
[1]   The role of automated urine particle flow cytometry in clinical practice [J].
Delanghe, JR ;
Kouri, TT ;
Huber, AR ;
Hannemann-Pohl, K ;
Guder, WG ;
Lun, A ;
Sinha, P ;
Stamminger, G ;
Beier, L .
CLINICA CHIMICA ACTA, 2000, 301 (1-2) :1-18
[2]   Microbiological evaluation of a commercial transport system for urine samples [J].
Eriksson, I ;
Lindman, R ;
Thore, M .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2002, 62 (05) :325-335
[3]  
European Confederation of Laboratory Medicine, 2000, SCAND J CLIN LAB I S, V231, P1, DOI DOI 10.1080/00365513.2000.12056993
[4]  
FRASER CG, 1989, CRIT REV CL LAB SCI, V27, P409
[5]  
Froom P, 2000, CLIN CHEM, V46, P1384
[6]  
Kouri T, 2002, CLIN CHEM, V48, P900
[7]   The urine test strip of the future [J].
Kutter, D .
CLINICA CHIMICA ACTA, 2000, 297 (1-2) :297-304
[8]   EVALUATION OF PRESERVATIVE FLUID FOR URINE COLLECTED FOR CULTURE [J].
LAUER, BA ;
RELLER, LB ;
MIRRETT, S .
JOURNAL OF CLINICAL MICROBIOLOGY, 1979, 10 (01) :42-45
[9]  
Penders J, 2002, CLIN CHEM, V48, P2236
[10]   BORIC ACID PRESERVATION OF URINE SAMPLES [J].
PORTER, IA ;
BRODIE, J .
BRITISH MEDICAL JOURNAL, 1969, 2 (5653) :353-&