The optimal amount of isotopic spike solution for ultratrace analysis by isotope dilution mass spectrometry

被引:30
作者
Hoelzl, R
Hoelzl, C
Kotz, L
Fabry, L
机构
[1] Wacker Siltron AG, D-84479 Burghausen, Germany
[2] Univ Regensburg, Inst Inorgan Chem, D-93040 Regensburg, Germany
[3] Univ Regensburg, Inst Phys Chem, D-93040 Regensburg, Germany
关键词
isotope dilution; mass spectrometry; ultratrace analysis; calibration; ultra-pure water;
D O I
10.1007/s007690050219
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Inductively coupled plasma mass spectrometry (ICP-MS) and high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) are powerful methods of determining metallic impurities in the low-and sub-ppt level in process media such as ultra-pure water used in semiconductor and wafer manufacturing. By using mass spectrometers for analysis, an isotope dilution analysis (IDMS) is possible. The reproducibility of an IDMS is unmatched. For concentration levels near the instrument detection limit a novel method is reported to find the optimal amount of isotopic spike solution. This optimal value can be derived by the law of propagation of uncertainty combined with the Poisson statistics of the measured number of counts. Generally! an excess of isotopic spike solution should be used to provide results of lowest possible uncertainty. The results are presented in a diagram for easy practical use.
引用
收藏
页码:185 / 188
页数:4
相关论文
共 7 条
[1]  
[Anonymous], 1988, INORGANIC MASS SPECT
[2]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[3]   ISOTOPE-DILUTION MASS-SPECTROMETRY - WHAT CAN IT CONTRIBUTE TO ACCURACY IN TRACE ANALYSIS [J].
DEBIEVRE, P .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (07) :766-771
[4]   PRECISION MASS SPECTROMETRIC ISOTOPE DILUTION ANALYSIS [J].
DEBIEVRE, PJ ;
DEBUS, GH .
NUCLEAR INSTRUMENTS & METHODS, 1965, 32 (02) :224-&
[5]  
Fabry L., 1996, ACCREDIT QUAL ASSUR, V1, P99
[6]   MASS-SPECTROMETRIC ISOTOPE DILUTION ANALYSIS OF SMALL CALCIUM CONCENTRATIONS IN MINERALS [J].
HEUMANN, KG ;
KUBASSEK, E ;
SCHWABENBAUER, W .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1977, 287 (2-3) :121-127
[7]  
*INT UN PUR APPL C, 1994, PURE APPL CHEM, V66, P2423