The bispectral index declines during neuromuscular block in fully awake persons

被引:188
作者
Messner, M
Beese, U
Romstöck, J
Dinkel, M
Tschaikowsky, K
机构
[1] Univ Erlangen Nurnberg, Dept Anesthesiol, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Dept Neurosurg, D-91054 Erlangen, Germany
[3] Frankenwaldklin, Dept Anesthesiol, Kronach, Germany
关键词
D O I
10.1213/01.ANE.0000072741.78244.C0
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Bispectral index (BIS) is an electroencephalographic variable promoted for measuring depth of anesthesia. Electromyographic activity influences surface electroencephalography and the calculation of BIS. In this study, we sought to determine the effect of spontaneous electromyographic activity on BIS. BIS was monitored in three volunteers by using an Aspect A-1000 monitor. The experiment was repeated in one volunteer. Electromyographic activity was recorded. Alcuronium and succinylcholine were administered. No other drugs were used. In parallel with spontaneous electromyographic activity of the facial muscles, BIS decreased in response to muscle relaxation to a minimum value of 33 and, in the repeated measurement, to a minimum value of 9 when total neuromuscular block was achieved. In two volunteers, no total block was achieved. BIS decreased to a minimal value of 64 and 57, respectively. In turn, recovery of BIS coincided with the reappearance of spontaneous electromyographic activity. During the entire experiment, the volunteers had full consciousness. BIS, assessed by software Version 3.31, correlates with spontaneous electromyographic activity of the facial muscles. BIS failed to detect awareness in completely paralyzed subjects. Thus, in paralyzed patients, BIS monitoring may not reliably indicate a decline in sedation and imminent awareness.
引用
收藏
页码:488 / 491
页数:4
相关论文
共 10 条
[1]   PLASMA-CONCENTRATIONS OF PANCURONIUM AND NEUROMUSCULAR BLOCKADE AFTER INJECTION INTO THE ISOLATED ARM, BOLUS INJECTION, AND CONTINUOUS INFUSION [J].
AGOSTON, S ;
FELDMAN, SA ;
MILLER, RD .
ANESTHESIOLOGY, 1979, 51 (02) :119-122
[2]   Electromyographic activity falsely elevates the bispectral index [J].
Bruhn, J ;
Bouillon, TW ;
Shafer, SL .
ANESTHESIOLOGY, 2000, 92 (05) :1485-1487
[3]   Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane, and alfentanil in healthy volunteers [J].
Glass, PS ;
Bloom, M ;
Kearse, L ;
Rosow, C ;
Sebel, P ;
Manberg, P .
ANESTHESIOLOGY, 1997, 86 (04) :836-847
[4]   Falsely increased bispectral index values in a series of patients undergoing cardiac surgery using forced-air-warming therapy of the head [J].
Hemmerling, TM ;
Fortier, JD .
ANESTHESIA AND ANALGESIA, 2002, 95 (02) :322-323
[5]   Development and clinical application of electroencephalographic bispectrum monitoring [J].
Johansen, JW ;
Sebel, PS .
ANESTHESIOLOGY, 2000, 93 (05) :1336-1344
[6]   Electroencephalographic derivatives as a tool for predicting the depth of sedation and anesthesia induced by sevoflurane [J].
Katoh, T ;
Suzuki, A ;
Ikeda, K .
ANESTHESIOLOGY, 1998, 88 (03) :642-650
[7]   Bispectral analysis of the electroencephalogram predicts conscious processing of information during propofol sedation and hypnosis [J].
Kearse, LA ;
Rosow, C ;
Zaslavsky, A ;
Connors, P ;
Dershwitz, M ;
Denman, W .
ANESTHESIOLOGY, 1998, 88 (01) :25-34
[8]   A primer for EEG signal processing in anesthesia [J].
Rampil, IJ .
ANESTHESIOLOGY, 1998, 89 (04) :980-1002
[9]   Continuous electromyography monitoring of motor cranial nerves during cerebellopontine angle surgery [J].
Romstöck, J ;
Strauss, C ;
Fahlbusch, R .
JOURNAL OF NEUROSURGERY, 2000, 93 (04) :586-593
[10]   A multicenter study of bispectral electroencephalogram analysis for monitoring anesthetic effect [J].
Sebel, PS ;
Lang, E ;
Rampil, IJ ;
White, PF ;
Cork, R ;
Jopling, M ;
Smith, NT ;
Glass, PSA ;
Manberg, P .
ANESTHESIA AND ANALGESIA, 1997, 84 (04) :891-899