Arterial input functions from MR phase imaging

被引:46
作者
Akbudak, E
Conturo, TE
机构
[1] UNIV WASHINGTON,SCH MED,MALLINCKRODT INST RADIOL,NEUROIMAGING LAB,DIV RADIOL SCI,ST LOUIS,MO 63110
[2] WASHINGTON UNIV,DEPT PHYS,ST LOUIS,MO 63130
关键词
MRI; perfusion; contrast agents; blood flow;
D O I
10.1002/mrm.1910360602
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method is presented for obtaining high-sensitivity arterial input functions following bolus intravenous contrast agent administration. Arterial contrast agent is monitored by phase reconstruction of single-shot echo-planar images. During bolus injections of a gadolinium (Gd) agent in a baboon, data were acquired at the mid-abdominal aorta, and magnitude and phase-shift images were reconstructed. Pairwise image subtraction was used to minimize phase aliasing. The phase-based method is shown to have a significant potential improvement in sensitivity compared to the magnitude approach. The phase method also has a general linear response to concentration. This method may have potential utility in quantitative imaging of blood flow and contrast agent kinetics.
引用
收藏
页码:809 / 815
页数:7
相关论文
共 38 条
[1]  
AKBUDAK E, P SMR 2 ANN M SAN FR, P938
[2]  
AKBUDAK E, P SMR 3 ANN M NIC FR, P867
[3]  
BERNE RM, 1967, CARDIOVASCULAR PHYSL, P78
[4]   FACTORS IN MYOCARDIAL PERFUSION IMAGING WITH ULTRAFAST MRI AND GD-DTPA ADMINISTRATION [J].
BURSTEIN, D ;
TARATUTA, E ;
MANNING, WJ .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :299-305
[5]   REAL-TIME MOVIE IMAGING FROM A SINGLE CARDIAC CYCLE BY NMR [J].
CHAPMAN, B ;
TURNER, R ;
ORDIDGE, RJ ;
DOYLE, M ;
CAWLEY, M ;
COXON, R ;
GLOVER, P ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (03) :246-254
[6]  
CHU SCK, 1990, MAGN RESON MED, V13, P239, DOI 10.1002/mrm.1910130207
[7]   SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS [J].
CONTURO, TE ;
SMITH, GD .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :420-437
[8]   MR IMAGING OF CEREBRAL PERFUSION BY PHASE-ANGLE RECONSTRUCTION OF BOLUS PARAMAGNETIC-INDUCED FREQUENCY-SHIFTS [J].
CONTURO, TE ;
BARKER, PB ;
MATHEWS, VP ;
MONSEIN, LH ;
BRYAN, RN .
MAGNETIC RESONANCE IN MEDICINE, 1992, 27 (02) :375-390
[9]  
CONTURO TE, 1993, J MAGN RESON IMAGING, V3, P62
[10]   IMPROVED SENSITIVITY TO MAGNETIC-SUSCEPTIBILITY CONTRAST [J].
DECRESPIGNY, AJ ;
ROBERTS, TPL ;
KUCHARCYZK, J ;
MOSELEY, ME .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (01) :135-137