Continuous dynamic-gravimetric preparation of calibration gas mixtures for air pollution measurements

被引:16
作者
Knopf, D [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
gas mixtures; mass flow; permeation tube; MFM; uncertainty determination;
D O I
10.1007/s007690000262
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The preparation of calibration gas mixtures for air pollution measurements by the dynamic-gravimetric method was investigated using sulphur dioxide in nitrogen as a model. The target mole fraction was 200 x 10(-9) mol/mol. with the option of also getting smaller mole fractions. Thermal mass now meters calibrated with reference mass flows were used to measure the dilution gas flow (nitrogen). The relative standard uncertainty of the dilution gas flows between 10 mg/s (approx. 500 ml/min) and 40 mg/s (approx. 2000 ml/min) was 0.15%. The mass now of the target component measured as the permeation rate was determined via the quasi-continuous observation of the loss in the permeation tube mass during the measuring time. A magnetic coupling system and an adapted microbalance were used for this purpose. The results presented show permeation rates measured over the lifetime of a tubular permeation source. The measurement cycles took between 3 days and 7 h at least. The relative standard uncertainty of the mixture composition did not exceed 2%. First comparisons with gas mixtures prepared by the static-gravimetric method show compatibility. The applicability of the system is not restricted to the SO2/N-2 mixture. It can also be used for preparing other gas mixtures in this field of application.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 1 条
[1]  
*ISO, 1993, GUID EXPR UNC MEAS G