Evaluation of a foot-to-foot impedance meter measuring extracellular fluid volume in addition to fat-free mass and fat tissue mass

被引:25
作者
Jaffrin, MY [1 ]
Kieffer, R [1 ]
Moreno, MV [1 ]
机构
[1] Univ Technol Compiegne, CNRS, UMR 6600, Biomech & Biomed Engn Lab, F-60206 Compiegne, France
关键词
body composition; extracellular volume; bioimpedance spectroscopy; foot-to-foot impedancemetry;
D O I
10.1016/j.nut.2004.11.016
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 [营养与食品卫生学];
摘要
Objectives: The first objective was to compare the accuracy of a foot-to-foot impedance meter with a multifrequency bioimpedance. for measurements of fat-free mass (FFM) and fat mass (FM) using dual energy X-ray absorptiometry (DXA), as reference. The second objective was to validate measurements of extracellular water resistance and volume by the foot-to-foot impedance meter, using multifrequency bioimpedance as reference. Methods: This investigation was carried out in 60 volunteers 18 to 71 y of age. Impedance meters were a Tefal Bodymaster Vision (foot-to-foot) that featured a square wave signal and a Xitron Hydra 4200 (5 to 1000 kHz) by using the bioimpedance spectroscopic method. Results: Bland-Altman tests showed that FFM differences between Tefal and DXA data were 1.98 +/- 3.09 kg in men and -0.08 +/- 2.98 kg in women. Total body water was measured by the Xitron, and FFM as measured with the Xitron was calculated as total body water divided by 0.732. Mean differences between Xitron-measured and DXA-measured FFM were 2.37 +/- 3.03 kg for men and 2.84 +/- 2.40 kg for women, indicating a systematic underestimation by the Xitron of intracellular volume. Extracellular water resistances measured by Tefal were in good agreement with those measured by Xitron with electrodes pasted under the subject's feet (mean difference 8.5 +/- 31 Omega). Extracellular water volumes were calculated from Tefal-measured extracellular water resistances by using a modified bioimpedance spectroscopic method and differed from those measured with Xitron by -0.03 +/- 0.66 L. Conclusion: Limits of agreement with DXA-measured FFM produced by the foot-to-foot impedance meter tested are too large for clinical measurements in individuals, but they are sufficient to assess FFM in groups of subjects and for home use. Our prototype was also capable of estimating extracellular water volume with a similar accuracy as multifrequency bioimpedance in normal subjects. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:815 / 824
页数:10
相关论文
共 18 条
[1]
BOULIER A, 2002, ACTA DIABETOL, V39, P142
[2]
Predicting body cell mass with bioimpedance by using theoretical methods: A technological review [J].
DeLorenzo, A ;
Andreoli, A ;
Matthie, J ;
Withers, P .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (05) :1542-1558
[3]
Human hydrometry:: comparison of multifrequency bioelectrical impedance with 2H2O and bromine dilution [J].
Ellis, KJ ;
Wong, WW .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 85 (03) :1056-1062
[4]
Extracellular and intracellular volume variations during postural change measured by segmental and wrist-ankle bioimpedance spectroscopy [J].
Fenech, M ;
Jaffrin, MY .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (01) :166-175
[5]
Hanai T., 1968, EMULSION SCI
[6]
Continuous monitoring of plasma, interstitial, and intracellular fluid volumes in dialyzed patients by bioimpedance and hematocrit measurements [J].
Jaffrin, MY ;
Fenech, M ;
de Fremont, JF ;
Tolani, M .
ASAIO JOURNAL, 2002, 48 (03) :326-333
[7]
Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model [J].
Jebb, SA ;
Cole, TJ ;
Doman, D ;
Murgatroyd, PR ;
Prentice, AM .
BRITISH JOURNAL OF NUTRITION, 2000, 83 (02) :115-122
[8]
Bipedal bioelectrical impedance analysis reproducibly estimates total body water in hemodialysis patients [J].
Kushner, RF ;
Roxe, DM .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2002, 39 (01) :154-158
[9]
Single prediction equation for bioelectrical impedance analysis in adults aged 20-94 years [J].
Kyle, UG ;
Genton, L ;
Karsegard, L ;
Slosman, DO ;
Pichard, C .
NUTRITION, 2001, 17 (03) :248-253
[10]
Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents [J].
Lazzer, S ;
Boirie, Y ;
Meyer, M ;
Vermorel, M .
BRITISH JOURNAL OF NUTRITION, 2003, 90 (05) :987-992