Feasibility of using ultrasound phased arrays for MRI monitored noninvasive surgery

被引:47
作者
Hynynen, K [1 ]
Chung, A [1 ]
Fjield, T [1 ]
Buchanan, M [1 ]
Daum, D [1 ]
Colucci, V [1 ]
Lopath, P [1 ]
Jolesz, F [1 ]
机构
[1] HARVARD UNIV,SCH MED,BOSTON,MA 02115
关键词
D O I
10.1109/58.542049
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this paper was to evaluate the in vivo feasibility of using phased arrays for MRI guided ultrasound surgery, Two different array concepts were investigated: a spherically curved concentric ring array to move the focus along the central iuds and a spherically curved 16 square element array to make the focus larger. Rabbit thigh muscles were exposed in vivo in a 1.5 T MRI scanner to evaluate the array performance. The results showed that both of the arrays performed as expected, and the focus could be moved and enlarged, In addition, adequate power could be delivered from the arrays to necrose in vivo muscle tissue in 10 s, This study was the first implementation of phased arrays for MRI guided ultrasound surgery, The results demonstrate that phased arrays have significant potential for noninvasive tissue coagulation.
引用
收藏
页码:1043 / 1053
页数:11
相关论文
共 32 条
[1]   TRANSIENT FIELDS OF CONCAVE ANNULAR ARRAYS [J].
ARDITI, M ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1981, 3 (01) :37-61
[2]   A TAPERED PHASED-ARRAY ULTRASOUND TRANSDUCER FOR HYPERTHERMIA TREATMENT [J].
BENKESER, PJ ;
FRIZZELL, LA ;
OCHELTREE, KB ;
CAIN, CA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1987, 34 (04) :446-453
[3]   DESIGN AND EXPERIMENTAL EVALUATION OF ARM INTRACAVITARY ULTRASOUND PHASED-ARRAY SYSTEM FOR HYPERTHERMIA [J].
BUCHANAN, MT ;
HYNYNEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (12) :1178-1187
[4]   CONCENTRIC-RING AND SECTOR-VORTEX PHASED-ARRAY APPLICATORS FOR ULTRASOUND HYPERTHERMIA [J].
CAIN, CA ;
UMEMURA, S .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :542-551
[5]  
CHUNG A, MAGN RES MED
[6]   FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION [J].
CLINE, HE ;
HYNYNEN, K ;
WATKINS, RD ;
ADAMS, WJ ;
SCHENCK, JF ;
ETTINGER, RH ;
FREUND, WR ;
VETRO, JP ;
JOLESZ, FA .
RADIOLOGY, 1995, 194 (03) :731-737
[7]   MR TEMPERATURE MAPPING OF FOCUSED ULTRASOUND SURGERY [J].
CLINE, HE ;
HYNYNEN, K ;
HARDY, CJ ;
WATKINS, RD ;
SCHENCK, JF ;
JOLESZ, FA .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) :628-636
[8]   THE EFFECT OF VARIOUS PHYSICAL PARAMETERS ON THE SIZE AND SHAPE OF NECROSED TISSUE VOLUME DURING ULTRASOUND SURGERY [J].
DAMIANOU, C ;
HYNYNEN, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (03) :1641-1649
[9]   FOCAL SPACING AND NEAR-FIELD HEATING DURING PULSED HIGH-TEMPERATURE ULTRASOUND THERAPY [J].
DAMIANOU, C ;
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :777-787
[10]   ONLINE MONITORING OF ULTRASONIC SURGERY WITH MR-IMAGING [J].
DARKAZANLI, A ;
HYNYNEN, K ;
UNGER, EC ;
SCHENCK, JF .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (03) :509-514